Skip to main content
Fig.5 | Journal of Nanobiotechnology

Fig.5

From: MXene-assisted organic electrochemical transistor biosensor with multiple spiral interdigitated electrodes for sensitive quantification of fPSA/tPSA

Fig.5

Clinics assessments for 49 patient biosamples. a Correlation analysis between reference values (Beckham DXI600) and measured value (via isMOECTs) for quantification of a fPSA and b tPSA. c The T-test discrimination for the ratio of fPSA/tPSA among all biosamples. d The comparisons of receiver operating characteristic between the chemiluminescence immunoassay (blue curve, Beckham DXI600) and the isMOECTs biosensors (black curve)