From: MOF derived core-shell CuO/C with temperature-controlled oxygen-vacancy for real time analysis of glucose
Sample
Concentration recorded from hexokinase method (mM)
Concentration calculated from biosensor (mM)
R.S.D. (%)
Recovery (%)
5
13.2
12.81
2.2
97
6
11.7
11.82
1.8
101
7
4.7
4.61
2.3
98
8
6.1
5.92
2.1