Skip to main content

Table 2 Precision analysis of CRP estimation by RAP.

From: A high sensitivity assay for the inflammatory marker C-Reactive protein employing acoustic biosensing

Sample

[CRP] g/mL Expected

[CRP] g/mL by hsELISA

[CRP] g/mL by RAP

RAP S.D

RAP %CV

A

0.1

0.092

0.3 (n = 3)

0.026

8.82

B

1.2

1.26

1.3 (n = 3)

0.04

3.1

C

3.61

4.35

3.89 (n = 3)

0.45

11.6

D

10.4

9.6

11.34 (n = 4)

0.88

7.75

E

34.8

44

47.22 (n = 5)

5.97

12.65

F

94.8

116

116.9 (n = 4)

7.93

6.78

  1. Horse sera spiked with human CRP and appropriately diluted was determined within and between runs using the direct sandwich format.