Table 3 The ratio of successful fishing force detection in HeLa cells at low temperature: analysis on the data set of Fig. 3

From: High efficiency penetration of antibody-immobilized nanoneedle thorough plasma membrane for in situ detection of cytoskeletal proteins in living cells

 

4 °C

37 °C

Detection of fishing forces over threshold* (n out of 30 trials)

14

9

Detection of fishing force over threshold* (%)

46.7

30.0

Average of fishing force over threshold (nN) ± SD

100.7 ± 69.1

58.8 ± 26.8

S/N**

9.0

4.4

  1. * Threshold equals to (average fishing force of MCF-7 cells as negative control at each temperature) + 4 × (standard deviation)
  2. ** Ratio of (average fishing force of HeLa)/(average fishing force of MCF-7)