Highly sensitive and accurate detection of C-reactive protein by CdSe/ZnS quantum dot-based fluorescence-linked immunosorbent assay

Abstract

Background

The conventional and widely used enzyme-linked immunosorbent assays (ELISA), due to specificity and high-sensitivity, were suitable in vitro diagnosis. But enzymes are vulnerable to the external conditions, and the complex operation steps limit its application. Semiconductor quantum dots have been successfully used in biological and medical research due to the high photoluminescence and high resistance to photobleaching. In this study, we have developed a novel quantum dot-labeled immunosorbent assay for rapid disease detection of C-reactive protein (CRP).

Results

The assay for the detection of CRP can provide a wide analytical range of 1.56–400 ng/mL with the limit of detection (LOD) = 0.46 ng/mL and the limit of quantification = 1.53 ng/mL. The precision of the assay has been confirmed for low coefficient of variation, less than 10% (intra-assay) and less than 15% (inter-assay). The accuracy of assay meets the requirements with the recoveries of 95.4–105.7%. Furthermore, clinical samples have been collected and used for correlation analysis between this FLISA and gold standard Roche immunoturbidimetry. It shows excellent accurate concordance and the correlation coefficient value (R) is as high as 0.989 (n = 34).

Conclusions

This in vitro quantum dot-based detection method offers a lower LOD and a wide liner detection range than ELISA. The total reaction time is only 50 min, which is much shorter than the commercialization ELISA (about 120 min). All of the results show that a convenient, sensitive, and accurate fluorescence-linked immunosorbent assay method has been well established for the detection of CRP samples. Therefore, this method has immense potential for the development of rapid and cost-effective in vitro diagnostic kits.

Background

Currently, enzymes as labels have been widely used in commercially quantitative immunoassays [1, 2], especially enzyme-linked immunosorbent assay (ELISA) [3, 4] is frequently used in in vitro diagnosis because of its high sensitivity and specificity [5, 6]. The signal collection in ELISA are based on a change in optical transmittance by a time-dependent enzymatic reaction, but enzymes are usually vulnerable to external conditions and very susceptible to matrix constituent [7]. So these complicated operation steps may limit its application. Compared to ELISA, a new approach called fluorescence-linked immunosorbent assay (FLISA) has been developed recently [8,9,10]. The emission of fluorescent signal as the analytical signal can be immediately recorded by the microplate reader, and it reduces the analysis time because of no chromogenic reaction steps of ELISA [11].

Due to the excellent optical characteristics [12, 13], Semiconductor quantum dots (QDs) are ideal fluorescent probes for advanced biosensor [14, 15], cellular imaging [16,17,18], cell labeling [19, 20], gene sequencing [21, 22], and chemical analysis [23, 24]. Thereafter, QD-based fluorescence probes begin to be introduced in the field of in vitro diagnosis [25,26,27]. Lin and co-workers first integrated QDs into lateral flow immunoassay (LFIA) and used to detect the concentration of nitrated ceruloplasmin [28]. We also adapted LFIA for the detection of human chorionic gonadotrophin (HCG) antigen by using CdSe/ZnS QDs microspheres as fluorescence probes successfully [29]. The quantitative result showed that the sensitivity of HCG antigen detection could reach 0.5 IU L−1, which was almost 20 times higher than traditional LFIA using tinctorial labels. Most recently, we have developed a sensitive quantitative LFIA to detect influenza A virus subtypes H5 and H9 simultaneously [30], the accuracy is as high as that of real-time PCR.

As highly environmental-stable, photo-stable, and fluorescence quantum yield materials, QDs are potentially to be used as new generation fluorescence labels in FLISA and can be widely used in in vitro diagnosis [2, 31,32,33]. Recent developed “green” synthesis of QDs has successfully lower the total cost and provides an opportunity to use QDs as affordable labeling materials [34]. Here in this paper, we have developed a novel quantum dots-based FLISA for disease detection. C-reactive protein (CRP), as an acute phase protein from liver cells, has been used clinically to monitor infection and autoimmune disorders [35, 36]. Especially, high-sensitivity C-reactive protein (hs-CRP), much higher sensitivity than CRP, has been regarded as an independent biomarker for cardiovascular disease (CVD) in clinic [37, 38].

This FLISA approach employs a novel water-soluble QDs fluorescent probe by using CdSe as core and ZnS as shell, instead of enzyme-linked antibody, and the process is shown in Scheme 1. The high quality CdSe/ZnS core/shell QDs were synthesized according to the “green” phosphine-free and low-cost synthesis method developed by our group [34]. First, CRP antigens were captured by the coating CRP antibodies on the microplate. Then, the labeled antibodies combined to antigens, and a mAb-Ag-mAb sandwich structure was formed after sufficient incubation time. The fluorescence intensity was proportional to the concentration of CRP antigens. The optimal reactive conditions have been found in this study, including the stability of QDs-mAb and optimal incubation time etc. As expected, this novel method has been proven that it can not only shorten the total reaction time, but also simplify the operation steps. And more important, the detection sensitivity and accuracy have been greatly improved. The further development of this immunoassay has immense potential for future convenient and cost-effective in vitro nano-medical diagnostic kits.

Scheme 1
scheme 1

A schematic illustration of FLISA procedure

Methods

Materials and instruments

Cadmium oxide (CdO, 99.99%), zinc oxide (ZnO, 99.99%, power), sulfur (S, 99.98%, power), selenium (Se, 99.99%, power), oleic acid (OA, 90%), 1-octadecene (ODE, 90%), and 2-(N-morpholino) ethanesulfonic acid (MES) were purchased from Aldrich. NaOH, HCl, NaCl, KCl, Na2CO3, NaHCO3, KH2PO4, Na2HPO4, H3BO3, Na2B4O7∙10H2O, Tris, Hepes, and Tween-20 were purchased from Shanghai Sangon Ltd (China). Bovine serum albumin (BSA) and calf serum were purchased from Sigma. 1-ethyl-3-(3-(dimethylamino) propyl) carbodiimide (EDC), N-Hydroxysulfosuccinimide (sulfo-NHS) and the microplates were purchased from Thermo Fisher Scientific (USA). Mouse anti C-reaction protein monoclonal antibody and CRP antigen were obtained from Abcam (USA). The fluorescence spectra were detected using SpectraMaxi3 (Molecular Devices, Sunnyvale, USA). Images of electrophoresis gels were taken using a gel imaging system (GenoSens1860, Shanghai, China). The sizes of QDs and QD-antibody probe were recorded using dynamic light scattering (Nano-ZS 90, Malvern Instruments, UK). Purified water (18.2 mΩ, Millipore USA) was used in all experiments.

Synthesis of hydrophobic CdSe/ZnS QDs and aqueous CdSe/ZnS QDs

The synthesis of CdSe/ZnS core/shell QDs was performed according to the previous literature from our group [34]. The PL and absorption spectra of the hydrophobic QDs are shown in Additional file 1: Figure S1. Amphiphilic oligomers (polymaleic acid n-hexadecanol ester, PMAH) were chosen as surfactants to prepare aqueous CdSe/ZnS QDs in this work [29, 39]. The hydrophobic chains of amphiphilic oligomers were used to combine the hydrophobic coatings on the CdSe/ZnS QDs and free carboxylic acid groups were available for further surface modification. The CdSe/ZnS QDs could be transferred into water by a phase transfer method in Additional file 1: Figure S2 (detailed operations in supporting information).

Preparation of QDs-antibody detecting probes

The QDs were conjugated with monoclonal antibody (shown in Scheme 2) according to the following protocol: first, 150 μL of PMAH-coated aqueous QDs (5 mg/mL) dissolved in 750 μL sodium borate buffer (5 mM, pH 7.2, BS buffer), then it was mixed with 50 μL of suflo-NHS (49 mg/mL in 5 mM BS buffer, pH 7.2) and 50 μL of EDC (17.36 mg/mL in 5 mM MES buffer, pH 5.5). The mixture was activated to react at 4 °C for 5 min under ultrasonic. Therewith, the superstratum clear-solution was abandoned after centrifuge separation at 20,000 rpm and 4 °C for 30 min. Second, 27.6 μL monoclonal antibodies (2.5 mg/mL) were added into the QDs with dissolving in 200 μL BS buffer (5 mM, pH 9.0) and incubated at 37 °C for 2 h. Afterwards, the mixture was blocked by 1% BSA solution and incubated at 37 °C for 30 min. Finally, the probe was washed by 5 mM BS buffer (pH 8.0). The QDs-mAb was stored in 50 μL Tirs solution (5 mM, pH 8.0) for future use.

Scheme 2
scheme 2

Formation of QD labeled antibody

Preparation of antibody coated fluorescence microplate

Primary antibody (the concentration of CRP monoclonal antibody was 1.8 mg/mL) in every well of microplate was diluted with a carbonate–bicarbonate buffer (50 mM, pH 9.6, CB buffer). Subsequently, the microplate was covered with sealing film and incubated at 4 °C for 24 h. In order to remove extra coating antibody, the microplate was washed three times with a wash buffer (0.05% Tween-20 in 10 mM PBS, pH 7.4). Then excess binding sites were blocked with 0.5% (w/v) BSA in 10 mM PBS (pH 7.4) for incubating overnight at 4 °C, this process ensured that all the available and remaining binding sides of the microplate wells were covered. The microplate was dried in a constant temperature humidity chamber for 24 h, then it was stored at 4 °C until use.

Optimization of reaction condition

In order to obtain the optimum conditions, various analytical conditions have been studied, including the screening QD–mAb probe diluted buffer (detailed content provided in Additional file 1: Figure S4), the optimization of reaction time, the coated primary monoclonal antibody concentration and the dilution ratio of the QDs-mAb probe. Then, under the optimization of reaction condition, an ideal regression coefficient for the standard curve has been obtained.

Quantitative detection of the CRP standard antigens by fluorescence-linked immunoassay

In every well of a 96-well microplate that contained coating antibody, 100 μL of the standard antigens, diluted to a series of concentrations with the sample buffer (10% calf serum (v/v) in 0.1 M PBS), were incubated at 37 °C for 30 min. Then the plate was followed by five washes with a wash buffer (PBS-T) at 1 min per wash. For subsequent operation, added 100 μL of the QDs-mAb probes diluted with the probe buffer (10% calf serum (v/v) in 0.1 M PBS) into each well were incubated at 37 °C for 30 min. After five washes with a wash buffer (PBS-T), the fluorescence intensity of each well in the plates was automatically read out by SpectraMax i3 when 450 nm was used as excitation wavelength.

Analysis of human serum samples

A total of 34 CRP human serum samples were obtained from Shenzhen Sixth People’s Hospital. Roche immunoturbidimetry assay was also utilized as a reference method to assess the accuracy of FLISA. Analysis samples were mixed with 1 μL human serum samples and 399 μL the sample buffer (the human serum samples were diluted 400 times). Each sample was tested three times respectively, and the mean value of the results was chosen.

Results and discussion

Characterization of QDs-mAb probes

Fluorescence spectra of aqueous CdSe/ZnS QDs and the QDs-mAb have been collected, the results are shown in Fig. 1a. Generally, the PL peak position of high quality CdSe/ZnS core/shell QDs is mainly determined by the band gap of CdSe, and it is not affected by its hydrodynamic sizes and ligands. Although the hydrodynamic size of QDs is increased after labeled with CRP antibodies (shown in Fig. 1b), the fluorescence peaks were approximately no changes on peak position and peak shape, and kept symmetrical in both aqueous QDs and QDs-mAb solutions. But the fluorescence intensity was declined to 64% on account of centrifuge separation during the coupling process.

Fig. 1
figure 1

Fluorescence spectra (a), dynamic light scattering (b), and agarose gel electrophoresis imaging (c) of the aqueous CdSe/ZnS QDs and QDs-mAb; and PL intensity of FLISA with different antigen concentrations (d)

In order to investigate the effect of conjugation on the size of QDs, the aqueous QDs and QDs-mAb were characterized by dynamic light scattering and agarose gel electrophoresis. The dynamic light scattering result clearly shows the narrow and uniform size distribution in Fig. 1b, and the hydrodynamic size of QDs increases from 50 to 120 nm after the conjugation with CRP antibodies, which can suggest the successful formation of QDs-mAb. The agarose gel electrophoresis (shown in Fig. 1c) indicates that the QDs-mAb probe moves slower than the aqueous QDs, due to the larger size of the QDs-mAb probe compared to the aqueous QDs, which means CRP antibodies have been successfully connected. Generally, zeta potential analysis can be used to evaluate that the stability of colloids, the value above +30 mV, or below −30 mV is believed to be stable in water. The zeta potential of the aqueous CdSe/ZnS QDs and QDs-mAb was measured and shown in Additional file 1: Figure S3. After the antibody conjugation with aqueous QDs, zeta potential is only changed from −46.8 to −40.0 mV, which is still stable in water. Figure 1d obviously illustrates that the fluorescence intensity increases with the increasing of antigens concentration (the red column diagram). It turned out that the aqueous QDs and CRP antibody protein maintained the original activity after the coupling process. The admirable stability of the aqueous QDs can be applied to coupling of other antibody protein as fluorescent probe.

The stabilities and optical properties of QDs-mAb probe were examined and compared in physiological relevant environmental conditions. The PL intensity values of QDs-mAb in acidic-to-alkaline pH environments were measured in following experiment (Fig. 2a). QDs-mAb was stable in a wide range of pH (4–12), the fluorescence intensity was able to be kept >75% even after 1 h preservation. The QDs-mAb was diluted into different buffers with 1 h preservation at 37 °C, the fluorescence intensities were measured (0, 10, 20, 30, 40, 50, 60 min, respectively). As shown in Fig. 2b, the effect of different buffers on the fluorescence stability of QDs-mAb show that the highest fluorescence intensity of the QDs-mAb could be kept in Tris buffer. Subsequently, the influence of fluorescence intensity with Tris buffer (pH 8.0) at different ionic strengths was studied (shown in Fig. 2c). It indicated that the fluorescence intensity was increased with the decreasing of the concentration. The QDs-mAb in Tris buffer (5 mM pH 8.0) showed the maximum fluorescence intensity, which was chosen as the storage buffer for the following experiments. The result shown in Fig. 2d illustrated that the storage stability was studied by measuring the fluorescence spectra of QDs-mAb in Tris buffer for a period of time (stored in 4 °C). The fluorescence intensity decreased to 86.7% after three months storage, which explicitly instructed the novel QDs-mAb could be applied to industrial production due to the excellent stability.

Fig. 2
figure 2

The PL intensity of QDs-mAb in various pH (a) and in different buffers (b). Fluorescence spectra of QDs-mAb in Tris buffer solutions with different ionic strengths (c). The PL intensity of QDs-mAb probes versus different storage time (d) in Tris buffer solution

Chessboard titration to determine the best proportion of coated antibody and label antibody

To confirm the optimum conditions of coating antibody concentration and the dilution ratio of QDs-mAb probe, we designed a chessboard titration experiment (Table 1). The coating antibody was diluted to different concentrations (10, 5, 1, 0.1 μg/mL) with CB buffer, and 100 μL of various coating antibody solutions were fixed in microplate. Subsequently added different concentration antigen (0, 1, 50 ng/mL) in corresponding wells and incubated for 30 min at 37 °C. The diluted ratio was 1:100, 1:200, 1:300, 1:400 (v/v) with the probe buffer of the QDs-mAb, followed by incubating for 30 min at 37 °C. The optimum condition was chosen from higher fluorescence intensity, and it is that the concentration of coating monoclonal antibody was 5 μg/mL when the diluted ratio of QDs-mAb was 1:100.

Table 1 Optimization of working concentration

Optimization of incubating time

In order to optimize the fluorescence developing time in detection, the fluorescence intensity was detected in different developing time. Here we changed the incubating time of QDs-mAb probe after the CRP antigen (50 ng/mL) was incubated for 30 min. The results shown in Fig. 3a and b indicate that the optimum fluorescence developing time was 20 min, when a dynamic balance was reached between the coating mAb-Ag and labeled-mAb reaction. However, the max fluorescence intensity at 50 min began to decrease because the label antibodies tend to be decomposed from the mAb-Ag-labeled mAb compounds. Thereof we selected 20 min as the optimum reaction because it shortened the reaction time under the premise of guarantee the fluorescence intensity. In a word, this quantum dot-labeled immunosorbent assay method controlled the reactive time within 50 min, which shortened the reaction time over 70 min comparing to the commercialization ELISA (the reactive time about 120 min).

Fig. 3
figure 3

Evolution of PL intensity in different fluorescence developing time (a) and optimal fluorescence developing time (b)

Calibration curves and linearity of analytical results

In the assay, standard CRP antigen (20 μg/mL) was diluted to 1.56, 3.125, 6.25, 12.5, 25, 50, 100, 200, and 400 ng/mL with the sample buffer, then followed to be incubated and washed by the washing buffer. The QDs-mAb probe was added into corresponding wells of microplate after dilution, the fluorescence intensity in every microplate well was measured. As shown in Fig. 4a, the fluorescence intensity was gradually increased with the increasing concentration of the standard antigen. The standard curve was constructed by plotting the fluorescence intensity (Y) against the CRP sample concentration (X) of analytes. Figure 4b showed that a good linear correlation was obtained with the concentration range from 1.56 to 400 ng/mL and the best fit for the calibration curve was Y = 0.778X + 4.372 with R2 = 0.992. The limits of detection (LOD) and quantification (LOQ) were the key parameters for evaluating detection method. Here we tested the black wells (the negative control samples) ten times at the same time, according to the provisions of International Union of Pure and Applied Chemistry (IUPAC), the LOD was calculated as three times the standard deviation (SD) to the slope of the min-concentration calibration plot (LOD = 3 SD/slope) and LOQ was calculated as ten times the same ratio (LOQ = 10 SD/slope). The final result showed that the novel CRP assay had a LOD was 0.46 ng/mL and a LOQ was 1.53 ng/mL.

Fig. 4
figure 4

Photoluminescence spectra from FLISA for determination of CRP antigen (a) and standard curves (b) (n = 3)

The developed immunoassay is superior to most of the current analysis techniques for detection of CRP in terms of highly sensitivity [40, 41]. The common clinical laboratory assays used to quantify CRP have a lower detection range in μg/mL, including latex agglutination or commercially available POCT methods. Recent new methodology-based immunoassays, such as sandwich ELISA, fluorescence sandwich immunoassay began to improve sensitivities in the range from μg/mL to ng/mL. Here this new developed FLISA method achieves ng/mL, at the same time, it offers a broader range. In addition, the QDs-mAb probes and the coated microplates could still keep good stability as long as 90 days and its fluorescence intensity maintain around 60% of the original value (shown in Additional file 1: Figure S5).

Precision and accuracy

It is well known that precision is an important indicator of measuring in vitro diagnosis, including intra-assay and extra-assay of imprecision. Three concentrations of quality control was 10, 50, 200 ng/mL, respectively using three batches of CRP reagent. By measuring each concentration five times, we calculated quality-control serum values of the mean, standard deviation, and coefficient of variation (CV). As shown in Table 2, the results met relevant provision of the test of biological products, intra-assay within CV <10%, and inter-assay within CV <15%. This method possessed excellent stability and repeatability on account of the good intra-assay and the extra-assay.

Table 2 Imprecision the intra-assay and the extra-assay

Recovery measurements were performed to assess the overall accuracy of the QD-based assay. The negative serum samples were spiked with a series of known standard CRP antigens for analysis, and the final concentrations covered the low, medium, and high-risk levels. The results shown in Table 3 indicated that all of the recovery rates were within the range of 95.4–105.7%, and the standard deviation was less than 15%. Due to the high accuracy and excellent repeatability, this FLISA method allowed for high-throughput detection on clinical research.

Table 3 Recoveries of different concentrations of antigen in human negative serum

Specificity of FLISA

The interference factors (immunoglobulin G, triglyceride, hemoglobin, and bilirubin) were selected as cross-reaction interference to evaluate the specificity of the FLISA method. As shown in Fig. 5a, the result showed that interference (10 mg/mL of immunoglobulin G, 20 mg/mL of triglyceride, 10 mg/mL of hemoglobin, and 0.35 mg/mL of bilirubin) did not interfere with the immunoassay, and validated that the interference factors did not have specific Ag–Ab combination, which suggested that the assay for a target analyte was not affected by the presence of other interference. The background fluorescence interference was caused by the autofluorescence of specific biomolecules in the serum, which leaded to appear the weak, nonspecific signals. However, it did not affect the accuracy of the CRP detection by introducing a reagent blank control consisting of the negative serum.

Fig. 5
figure 5

Cross-reactionof FLISA (a), and comparison of results for CRP concentrations by FLISA and Roche immunoturbidimetry (b)

Clinical sample detection and analysis

In this study, we also collected thirty-four CRP human serum samples from Shenzhen Six People’s Hospital for real clinical sample test. The immunoturbidimetry is based on the precipitation reactions of Ag–Ab, the turbidity appears by polyethylene glycol (PEG) and is measured by optical instruments. Due to the good sensitivity and accuracy, the immunoturbidimetry is mostly used for clinical diagnosis method of CRP by many hospitals worldwide. And Roche immunoturbidimetry generally has been recognized as gold standard test in vitro diagnostic benchmark and it is used to refer to the most accurate test possible without restrictions. By comparing the results from Roche immunoturbidimetry and this FLISA, the linear regression equation is Y = 1.12X−2.56 and the highly significant correlation is 0.989 (>0.9) as shown in Fig. 5b, which indicates the FLISA method has excellent concordance as the gold Roche immunoturbidimetry. Therefore, this new established FLISA is able to rapidly analyze serum samples from clinical patients after appropriate dilution with a high sensitivity and accuracy.

Conclusion

Currently, the majority common assays do not provide high-throughput quantification of multiple analysis and simple operation steps. Therefore, the development of methods for the accurate, sensitive, rapid, and convenient detection is important need in vitro diagnosis. In this report, a simple, rapid, and high sensitivity method by a novel quantum dots-labeled immunosorbent assay has been developed, which only takes half of the analysis time and offers a lower LOD, a wide liner detection range than those of ELISA. The water-soluble QDs as novel fluorescent probe possess excellent stability, without influencing on the performance of antibody protein, and can be applied to combine other antibody or antigen protein. This novel analytical tool can play an important role in preliminary test of CRP protein levels, and provide wide application in clinical research because the assay meets the needs felicitously of the simplicity, sensitivity, and high-throughput with a short time and low cost. Moreover, the FLISA method gives a very promising outlook for lower LOD as well as high-throughput capability which shows the potential for in vitro diagnostic tool in the future.

Abbreviations

Ab:

antibody

Ag:

antigen

BS:

sodium borate buffer

BSA:

bovine serum albumin

CB:

carbonate-bicarbonate buffer

CRP:

C-reactive protein

CV:

coefficient of variation

CVD:

cardiovascular disease

DLS:

dynamic light scattering

ELISA:

enzyme-linked immunosorbent assay

EDC:

1-ethyl-3-(3-(dimethylamino) propyl) carbodiimide

FLISA:

fluorescence-linked immunosorbent assay

HCG:

human chorionic gonadotrophin

IUPAC:

International Union of Pure and Applied Chemistry

LFIA:

lateral flow immunoassay

LOD:

the limit of detection

LOQ:

the limit of quantification

mAb:

monoclonal antibody

MES:

2-(N-morpholino) ethanesulfonic acid

ODE:

1-octadecene

PBS:

phosphate buffered saline

PBS-T:

0.01 M PBS contains 0.05%(v/v) Tween-20

PEG:

polyethylene glycol

PL:

photoluminescence

PMAH:

polymaleic acid n-hexadecanol ester

POCT:

point-of-care testing

QDs:

quantum dots

QDs-mAb:

quantum dots labeled monoclonal antibody

R:

the correlation coefficient value

SD:

standard deviation

sulfo-NHS:

N-hydroxysulfosuccinimide

References

  1. Cho JH, Paek EH, Cho IH, Paek SH. An enzyme immunoanalytical system based on sequential cross-flow chromatography. Anal Chem. 2005;77:4091–7.

    Article  CAS  Google Scholar 

  2. Speranskaya ES, Beloglazova NV, Abe S, Aubert T, Smet PF, Poelman D, Goryacheva IY, Saeger D, Hens Z. Hydrophilic, bright, CulnS2 quantum dots as Cd-free fluorescent labels in quantitative immunoassay. Langmuir. 2014;30:7567–75.

    Article  CAS  Google Scholar 

  3. Engvall E, Perlmann P. Enzyme-linked immunosorbent assay (ELISA) quantitative assay of immunoglobulin G. Immunochemistry. 1971;8:871–4.

    Article  CAS  Google Scholar 

  4. Zhang HF, Ma XM, Hu SS, Lin Y, Guo LH, Qi B, Lin ZY, Chen GN. Highly sensitive visual detection of Avian Influenza A (H7N9) virus based on the enzyme-induced metallization. Biosens Bioelectron. 2016;79:874–80.

    Article  CAS  Google Scholar 

  5. Rifai N, Tracy RP, Ridker PM. Clinical efficacy of an automated high-sensitivity C-reactive protein assay. Clin Chem. 1999;45:2136–41.

    CAS  Google Scholar 

  6. Dixit CK, Vashist SK, MacCraith BD, O’Kennedy K. Multisubstrate compatible ELISA procedures for rapid and high sensitivity immunoassays. Nat Protoc. 2011;6:439–45.

    Article  CAS  Google Scholar 

  7. Beloglazova NV, Goryacheva LY, Niessner R, Knopp D. A comparison of horseradish peroxidase, gold nanoparticles and quantum dots as labels in non-instrumental gel-based immunoassay. Microchim Acta. 2011;175:361–7.

    Article  CAS  Google Scholar 

  8. Chen J, Xu F, Jiang HY, Hou YL, Rao QX, Guo PJ, Ding SY. A novel quantum dot-based fluoroimmunoassay method for detection of Enrofloxacin residue in chicken muscle tissue. Food Chem. 2009;113:1197–201.

    Article  CAS  Google Scholar 

  9. Zhu XL, Chen LL, Shen P, Jia JW, Zhang DB, Yang LT. High sensitive detection of Cry1Ab protein using a quantum dot-based fluorescence-linked immunosorbent assay. J Agric Food Chem. 2011;59:2184–9.

    Article  CAS  Google Scholar 

  10. Song EQ, Yu MQ, Wang YY, Hu WH, Cheng D, Swihart MT, Song Y. Multi-color quantum dot-based fluorescence immunoassay array for simultaneous visual detection of multiple antibiotic residues in milk. Biosens Bioelectron. 2015;72:320–5.

    Article  CAS  Google Scholar 

  11. Beloglazova NV, Speranskaya ES, De Saeger S, Hens Z, Abe S, Goryacheva IY. Quantum dot based rapid tests for zearalenone detection. Anal Bioanal Chem. 2012;403:3013–24.

    Article  CAS  Google Scholar 

  12. Smith AM, Duan H, Mohs AM, Nie SM. Bioconjugated quantum dots for in vivo molecular and cellular imaging. Adv Drug Deliv Rev. 2008;60:1226–40.

    Article  CAS  Google Scholar 

  13. Hildebrandt N. Biofunctional quantum dots: controlled conjugation for multiplexed biosensors. ACS Nano. 2011;5:5286–90.

    Article  CAS  Google Scholar 

  14. Marin S, Merkoci A. Direct electrochemical stripping detection of cystic-fibrosis-related DNA linked through cadmium sulfide quantum dots. Nanotechnology. 2009;20:055101.

    Article  Google Scholar 

  15. Algar WR, Wegner D, Huston AL, Blanco-Canosa JB, Stewart MH, Armstrong A, Dawson PE, Hildebrandt N, Medintz IL. Quantum dots as simultaneous acceptors and donors in time-gated forster resonance energy transfer relays: characterization and biosensing. J Am Chem Soc. 2012;134:1876–91.

    Article  CAS  Google Scholar 

  16. Laura TA, Jose MCF, Rosario P, Alfredo SM. Development of a quantum dot-based fluorescent immunoassay for progesterone determination in bovine milk. Biosens Bioelectron. 2011;26:4753–9.

    Article  Google Scholar 

  17. Jaiswal JK, Mattoussi H, Mauro JM, Simon SM. Long-term multiple color imaging of live cells using quantum dot bioconjugates. Nat Biotechnol. 2003;21:47–51.

    Article  CAS  Google Scholar 

  18. Lu ZS, Zhu ZH, Zheng XT, Qiao Y, Guo J, Li CM. Biocompatible fluorescence enhanced ZrO2-CdTe quantum dot nanocomposite for in vitro cell imaging. Nanotechnology. 2011;22:155604.

    Article  Google Scholar 

  19. Wu XY, Liu HJ, Liu JQ, Haley KN, Treadway JA, Larson JP, Ge NF, Peale F, Bruchez MP. Immunofluorescent labeling of cancer marker Her2 and other cellular targets with semiconductor quantum dots. Nat Biotechnol. 2003;21:41–6.

    Article  CAS  Google Scholar 

  20. Alivisatos AP, Gu WW, Larabell C. Quantum dots as cellular probes. Annu Rev Biomed Eng. 2005;7:55–76.

    Article  CAS  Google Scholar 

  21. Han MY, Gao XH, Su J, Nie SM. Quantum-dot-tagged microbeads for multiplexed optical coding of biomolecules. Nat Biotechnol. 2001;19:631–5.

    Article  CAS  Google Scholar 

  22. Derfus AM, Chen AA, Min DH, Ruoslahti E, Bhatia SN. Targeted quantum dot conjugates for siRNA delivery. Bioconjugate Chem. 2007;18:1391–6.

    Article  CAS  Google Scholar 

  23. Gill R, Zayats M, Willner I. Semiconductor quantum dots for bioanalysis. Angew Chem Int Ed. 2008;47:7602–25.

    Article  CAS  Google Scholar 

  24. Pinwattana K, Wang J, Lin CT, Wu H, Du D, Lin YH. CdSe/ZnS quantum dots based electrochemical immunoassay for the detection of phosphorylated bovine serum albumin. Biosens Bioelectron. 2010;26:1109–13.

    Article  CAS  Google Scholar 

  25. Zou ZX, Du D, Wang J, Smith JN, Timchalk C, Li YQ, Lin YH. Quantum dot based immunochromatographic fluorescent biosensor for biomonitoring trichloropyridinol, a biomarker of exposure to chlorpyrifos. Anal Chem. 2010;82:5125–33.

    Article  CAS  Google Scholar 

  26. Shen HB, Yuan H, Wu F, Bai XW, Zhou CH, Wang HZ, Lu TK, Qin ZF, Ma L, Li LS. Facile synthesis of high-quality CuInZnxS2+x core/shell nanocrystals and their application for detection of C-reactive protein. J Mater Chem. 2012;22:18632.

    Google Scholar 

  27. Liu D, Wu F, Zhou CH, Shen HB, Yuan H, Du ZL, Ma L, Li LS. Multiplexed immunoassay biosensor for the detection of serum biomarkers —B-HCG and AFP of down syndrome based on photoluminescent water-soluble CdSe/ZnS quantum dots. Sens Actuator B Chem. 2013;186:235–43.

    Article  CAS  Google Scholar 

  28. Li ZH, Wang Y, Wang J, Tang ZW, Pounds JG, Lin YH. Rapid and sensitive detection of protein biomarker using a portable fluorescence biosensor based on quantum dots and a lateral flow test strip. Anal Chem. 2010;82:7008–14.

    Article  CAS  Google Scholar 

  29. Zhou CH, Yuan H, Shen HB, Guo Y, Li XM, Liu D, Xu L, Ma L, Li LS. Synthesis of size-tunable photoluminescent aqueous CdSe/ZnS microspheres via a phase transfer method with amphiphilic oligomer and their application for detection of HCG antigen. J Mater Chem. 2011;21:7393–400.

    Article  CAS  Google Scholar 

  30. Wu F, Yuan H, Zhou CH, Mao M, Liu Q, Shen HB, Cen Y, Qin ZF, Ma L, Li LS. Multiplexed detection of influenza A virus subtype H5 and H9 via quantum dot-based immunoassay. Biosens Bioelectron. 2016;77:464–70.

    Article  CAS  Google Scholar 

  31. Larson DR, Zipfel WR, Williams RM, Clark SW, Bruchez MP, Wise FW, Webb WW. Water-soluble quantum dots for multiphoton fluorescence imaging in vivo. Science. 2003;300:1434–6.

    Article  CAS  Google Scholar 

  32. Luo Y, Zhang B, Jiang TL, Zhou DY, Huang JF, Fu WL. Sensitive and rapid quantification of C-reactive protein using quantum dot-labeled microplate immunoassay. J Transl Med. 2012;10:24.

    Article  CAS  Google Scholar 

  33. Duan H, Chen XL, Xu W, Fu JH, Xiong YH, Wang A. Quantuam-DoT submicrobead based immunochromatographic assay for quantitative and sensitive detection of zearalenone. Talanta. 2015;132:126–31.

    Article  CAS  Google Scholar 

  34. Shen HB, Wang HZ, Tang ZJ, Niu JZ, Lou SY, Du ZL, Li LS. High quality synthesis of monodisperse zinc-blende CdSe and CdSe/ZnS nanocrystals with a phosphine-free method. Cryst Eng Comm. 2009;11:1733–8.

    Article  CAS  Google Scholar 

  35. Du Clos TW, Mold C. C-reactive protein—An activator of innate immunity and a modulator of adaptive immunity. Immunol Res. 2004;30:261–77.

    Article  Google Scholar 

  36. Yeh ETH, Willerson JT. Coming of age of C-reactive protein using inflammation markers in cardiology. Circulation. 2003;107:370–2.

    Article  Google Scholar 

  37. Out D, Hall RJ, Grangar DA. Assessing salivary C-reactive protein: longitudinal associations with systemic inflammation and cardiovascular disease risk in women exposed to intimate partner violence. Brain Behav Immun. 2012;26:543–51.

    Article  CAS  Google Scholar 

  38. Yang Z, Liu Y, Lei C, Sun XC, Zhou Y. A flexible giant magnetoimpedance-based biosensor for the determination of the biomarker C-reactive protein. Microchim Acta. 2015;182:2411–7.

    Article  CAS  Google Scholar 

  39. Zhou CH, Shen HB, Guo L, Xu L, Niu JZ, Zhang ZJ, Du ZL, Chen JM, Li LS. A versatile method for the preparation of water-soluble amphiphilic oligomer-coated semiconductor quantum dots with high fluorescence and stability. J Colloid Interface Sci. 2010;244:279–85.

    Article  Google Scholar 

  40. Algarra M, Campos BB, Gomes D, Alonso B, Casado CM, Arrebola MM. Thiolated DAB dendrimer/ZnSe nanoparticles for C-reactive protein recognition in human serum. Talanta. 2012;99:574–9.

    Article  CAS  Google Scholar 

  41. Vashist SK, Venkatesh AG, Schneider EM, Beaudoin C, Luppa PB, Luong JHT. Bioanalytical advances in assays for C-reactive protein. Biotechnol Adv. 2015;34:272–90.

    Article  Google Scholar 

Download references

Authors’ contributions

YL participated in the design of the study, optimized the reaction conditions for in vivo detection of CRP, performed the statistical analysis and drafted the finalized the manuscript. RW and LSL explained the experimental phenomena and modified the manuscript. KF provide guidance on the study. JL made hydrophobic CdSe/ZnS QDs to surface modification. QM prepared the QDs-antibody detecting probes. HY shared experiences in preparation of the QDs-antibody experiments. HS provided the hydrophobic CdSe/ZnS QDs. XC initiated the study, supervised the experiments and data interpretations. All authors read and approved the final manuscript.

Acknowledgements

The authors gratefully acknowledge the financial support from the research project of 113 Talents Program of China Medical City, Program for High-Level Innovation and Entrepreneurship of Jiangsu Province, Program for Science &Technology Innovation Talents in University of Henan Province (No. 14HASTIT009), and National Natural Science Foundation of China (21671058 and 61474037).

Competing interests

The authors declare that they have no competing interests.

Funding

National Natural Science Foundation of China.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding authors

Correspondence to Ruili Wu, Xiangdong Chai or Lin Song Li.

Additional file

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lv, Y., Wu, R., Feng, K. et al. Highly sensitive and accurate detection of C-reactive protein by CdSe/ZnS quantum dot-based fluorescence-linked immunosorbent assay. J Nanobiotechnol 15, 35 (2017). https://doi.org/10.1186/s12951-017-0267-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12951-017-0267-4

Keywords