Impact of nano-morphology, lattice defects and conductivity on the performance of graphene based electrochemical biosensors

Abstract

Diverse properties of graphenic materials have been extensively explored to determine properties that make good electrochemical nanomaterial-based biosensors. These are reviewed by critically examining the influence of graphene nano-morphology, lattice defects and conductivity. Stability, reproducibility and fabrication are discussed together with sensitivity and selectivity. We provide an outlook on future directions for building efficient electrochemical biosensors.

Highlights

  • Fabrication of electrochemical biosensors

  • Influence of nano-morphology, lattice defects and conductivity

  • Concepts and parameters governing electrochemical biosensor sensitivity, selectivity, stability and reproducibility

Introduction

Biosensors, broadly applicable to biology and biomedicine, can transform healthcare with innovative precise detection of scarce analytes via a biorecognition element and a transducer [1, 2]. Specific detection between an analyte (e.g. protein, enzyme, nucleic acid, biomarker molecule) and corresponding aptamer (apt) bioelement is transformed into a measurable signal by a transducer defining the biosensor type (e.g. electrical, electrochemical, optical, thermal, piezoelectrical or magnetic) [3]. A desirable biosensor characteristic is real-time quantitative analyte concentration measurement within a complex environment [4]. Since Leland Clark’s 1962 vanguard biosensor, designs have advanced considerably. Advantages sought include high sensitivity and selectivity, low-cost, simplicity, rapid response, low background noise with a strong signal-to-noise ratio (S/N), allowing label-free and environmentally friendly techniques [5, 6]. Nanotechnology has introduced many advanced materials, such as quantum dots (QDs), carbon nanotubes (CNT), and lately two-dimensional materials such as graphene. With peak surface-to-volume ratios, 2D materials introduce interfacial phenomena that can greatly improve biosensor sensitivity, selectivity, response time, and limits of detection [7, 8]. Graphene may be considered a biosensor material of choice [9] and a benchmark for future electrochemical sensor nanomaterials, such as boron nitride, transition metal dichalcogenides (TMDs), black phosphorus (BP), silicene and antimonene [6, 7]. The extensive scientific exploration of graphene and TMDs has been comprehensively reviewed [10]. Lately, composite nanomaterials of more complex architecture have sought to improve electron transfer qualities and analyte sensitivity, however, their fabrication may differ between laboratories and subtle alterations in characteristics may confusingly influence properties and performance.

Here, we introduce concepts for the main techniques used in electrochemical biosensing, highlighting the distinguishing qualities of advanced nanomaterials used for electrochemical biosensor elaboration. For graphenic sensors in particular, we focus on three key parameters: morphology, electrical conductivity and the influence of lattice defects.

Techniques used in electrochemistry

Electrochemical sensors represent promising and robust analytical tools, increasingly used in many fields, including analytical chemistry, molecular biology, healthcare diagnostics, environmental monitoring and security [2]. Electrochemical sensors generate electrical signals proportional to an analyte concentration. Conventionally, electrochemical events are often measured using a cell with three electrodes connected to a potentiostat: a working electrode (WE), a reference electrode (RE) and an auxiliary counter electrode (CE). Electrochemical processes involve charge transfer across the interface of the WE and an electrolyte. The three-electrode configuration allows measurement of a potential (current) between RE (CE) and WE with minimal ohmic drop interference. The current flowing through the RE is minimised to avoid its polarisation, thus stabilising the potential between the WE and RE.

Different electrochemical sensing technique nomenclature reflects what is measured: (1) Amperometric, current at a fixed potential; Voltammetric, current at a varying potential. (2) Potentiometric, potential or charge accumulation at constant current. (3) Conductometric, WE conductivity at varying frequencies. (4) Impedimetric, impedance (both resistance and reactance) at varying AC (alternating current) potential frequencies. (5) Field-effect, current generated by a potentiometric effect at the gate electrode using a transistor [11]. Typical voltammetry methods used in electrochemical biosensing include cyclic voltammetry (CV) [12], differential pulse voltammetry (DPV), and square wave voltammetry (SWV). Although electrochemical techniques are versatile, amperometry, voltammetry and electrochemical impedance spectroscopy (EIS) are favored in microfluidics and biosensing application [13].

In amperometry, the current (I) is recorded as a function of time (t) and various amperometric based sensors have been designed to measure the current continuouly from the oxidation or reduction of an electroactive species in an electrochemical reaction. The defining Clark experiment introduced the simplest form of amperometric biosensor, whereby currents produced by consumption of oxygen or production of hydrogen peroxide by oxydoreduction activities of glucose oxidase (GOx) permitted measurement of glucose concentration. Nowadays, amperometric biosensors are commercially available and widely used for glucose monitoring, reflecting the pioneering design’s effectiveness. Exemplifying current design progress, a miniaturised screen printed biosensor (100–400 microns), using a water-based ink containing cobalt phthalocyanine, detected glucose by chronoamperometry (CA) in the linear range from 0.5 to 2.5 mM [14], appropriate for cell toxicity applications.

Voltammetry, a leading electrochemistry technique, contrasts amperometry by being able to probe the reversibility of the studied system, since the electrode potential (E) changes as a function of time. Notably, [Fe(CN)6]3−/4− is a standard electrolyte for electrochemical biosensing, because of higher surface sensitivity in comparison to other redox systems such as [Ru(NH3)6]3+/2+ [15]. In cyclic voltammetry, once the setting potential is reached, the working electrode’s potential is inversely ramped to return to the initial potential. Compared to other electrochemical methods, CV simply and quickly evaluates the working electrode efficiency, both qualitatively and moreover quantitatively via several useful parameters: (i) Surface area, using Randles–Sevcik equation; (ii) Reversibility of the electrochemical process, estimating the redox peak-to-peak potential separation difference (ΔE) (ΔE should be minimised, for one-electron process ΔE is 56.9 mV in theory); (iii) Heterogenous electron transfer rate constant (k° in cm s−1) using Nicholson equations [16, 17]; and (iv) Electrode transfer rate (Ks in s−1) from the Laviron equation [18]. Those parameters crucially determine the possible electron transfer speed and how it could influence biosensor sensitivity. Independent of the scan rate and electrode surface, the shape and current intensity of the CV curve change with analyte concentration. In DPV, the current is determined directly before each potential change (pulse period) and the current difference is represented as a function of the potential. DPV usefully eliminates the contribution of non-Faradaic (capacitive) processes, enhancing the precision of electrode reaction analysis, whilst minimizing the charging current for high sensitivity.

Various sensors based on CV and DPV transducer signal have been described [19]. Recently, CV and DPV were used to detect estriol (ET), one of four ovarian estrogens strongly influencing sexual and reproductive function. For CV curves, peak current increased linearly from 2 × 10−6 to 1 × 10−4 M ET and a detection limit of 8.7 × 10−7 M was achieved [20]. Innovatively, DPV was used to sense salbutamol sulfate (SBS), a bronchodilator for asthma treatment banned by anti-doping agencies due to ergogenic action and side-effects including tachycardia and arrhythmia at high doses. The peak current response to SBS concentration covered a 0.2 to 8 µM linear range with a detection limit of 6.8 × 10−8 M [21]. Chlorpromazine, used to treat depression, bipolar disorder and schizophrenia, can have serious overdose side-effects, including interpalpebral conjunctiva and cataracts. A suitable DPV sensor could monitor chlorpromazine doses in the range of 0.01 to 0.08 µM with a detection limit of 0.003 µM [22].

Electrochemical impedance spectroscopy, sometimes called AC impedance or impedance spectroscopy is a non-destructive and powerful technique providing time dependent quantitative data regarding the electrode processes and complex interfaces. Its growing popularity for biomolecule detection reflects a high interface binding event sensitivity. In EIS, both a DC (direct current) potential and a small sinusoidal AC perturbation potential (EAC ~ 5–10 mV) are applied between the WE and the RE. The sinusoidal perturbations of the potential E(t) produces a sinusoidal current I(t) of the same frequency (ω) but shifted with a phase ϕ with respect to the potential. The magnitude |Z| and the phase angle ϕ of the recorded complex impedance reflects a function of the AC frequency after calculation using Ohm’s law (Z = V/I = |Z|e = Zreal + jZimag). The real part Zreal is similar to a resistance R (ϕ = 0, independent of frequency) while the imaginary part (Zimag) is the reactance (ϕ ≠ 0, dependent of frequency). In electrochemical biosensing, EIS is usually presented as a Nyquist plot, i.e. the dependence of Zimag as a function of Zreal. The diameter of the semi-circle in the middle frequency region of a Nyquist plot, representing charge transfer resistance (Rct), is related to the analyte concentration. Since the reactance is generally capacitive, i.e. of negative value,—Zimag is always plotted for convenience. The dependence of |Z| and ϕ as function of AC frequency, respectively known as Bode amplitude and Bode phase plots are rarely described, yet they can provide additional relevant information. For quantitative analysis, Nyquist and Bode diagrams are usually modeled using Randles circuit [23]. Charge transfer resistance responsiveness to analyte concentrations yielded sensitive impedance biosensors [24]. For example, the impedimetric label-free aptamer biosensor for lysozyme detection showed a limit of detection (LOD) of 1.67 µM, a linear response up to 5 µM, and a sensitivity of 0.090 µM−1 in relative charge transfer resistance values [25]. Electrochemical transducers provide an attractive means of converting a biological event to an electric signal. Recently, more advanced biosensors have been fabricated coupling electrochemical techniques and nanotechnology.

Toward the use of graphenic nanomaterials in biosensing

Standard carbonaceous electrodes

Numerous reviews highlight use of carbon versus mercury, platinum or gold for advanced electrochemical electrodes, since carbon materials are relatively non-toxic, chemically inert, cost-effective, having a wide potential window and provide good stability. Highly biocompatible, carbon facilitates covalent anchoring of specific biological species such as enzymes, proteins and DNA molecules. Glassy carbon electrode (GCE), pencil graphite electrode (PGE) and screen printed carbon electrodes (SPCE) are commonly used biosensor electrodes [26, 27].

GCE is a non-graphitic carbon, combining glassy and ceramic properties with those of graphite, synthesized by high temperatures pyrolysis of certain polymeric precursor above 2000 °C. GCE is chemically stable and compared to other types of carbon materials it possesses rather low reactivity, low oxidation rate, high chemical inertness, high hardness, impermeability due to very small pore sizes, and good electrical conductivity which make it a competitive inert, conductive electrode.

With customizable conductive inks and three-dimensional (3D) printing technology, more reproducible than simpler drop-casting methods, SPCE benefit from simplicity, cost-effectiveness, and suitability for mass production. Also, SPCEs enable simple integration and desired portability [28]. PGEs, although more fragile than SPCEs, are also a readily available and disposable low-cost carbon-based device. Screen printed pencil graphite electrodes can be conveniently designed in a similar shape to SPCE [29]. However, establishing clean electrodes before any surface modification is particularly important because they are subject to unintentional adsorption of impurities.

Biological compound detection with bare electrodes is difficult due to poor responsiveness and high over-potentials. Hence, electrode surface modification seeks to improve the electrochemical characteristics, either by pretreatment [30] or by chemical modification (e.g. electrooxidation or electroreduction). Prasad et al. pretreated bare-SPCE by applying a potential at 2.0 V for 300 s in pH 7.4 PBS (phosphate buffer solution) when investigating oxygen functional groups and edge plane sites on SPCE for the determination of dopamine (DA), uric (UA) and ascorbic (AA) acids [31]. The edge plane sites were the principal location for electron transfer arising from oxygen functionalities on active sites. Electrode pretreatment could introduce more edge planes and oxygen groups in the lattice. Unlike bare-SPCE or oxygen plasma treated SPCE, electrochemically pretreated SPCE* could simultaneously detect DA, UA and AA [31]. Edge plane sites on the surface improved electron transfer kinetics and increased resistance to surface passivation. In contrast to SPCE and PGE, bare-GCE contained a significant number of edge plane sites, confirmed by Raman spectroscopy. However, bare-GCE have low oxygen content and despite the introduction of oxygen functionalities by treatment in acidic solution [27] or plasma [32], the efficiency of such modified GCE for electrochemical biosensors remained debatable. Electrode topography also directly influenced analyte detection sensitivity. For example, high electrochemical reactivity of PGE, from an irregular surface morphology, increased the active surface area of the electrode (e.g., 0.255 cm2 for PGE compared to 0.0951 cm2 for carbon paste electrode) [26]. In the design of an electrochemical label-free biosensor detecting microRNA-125a, Yammouri et al. used PGE modified with several carbon nanomaterials, including carbon black (CB), multiwalled carbon nanotubes (MWCNT) and graphene oxide (GO). Their lowest LOD was 10 pM (1 pg/mL) obtained using PGE modified with CB providing a linear range between 1 nM and 2 μM [33]. Exploratory use of pretreated bare carbonaceous electrodes in electrochemical biosensors for healthcare applications introduced key questions; could we detect molecules with high efficiency on surface-modified GCE, PGE or SPCE and would this surpass biosensors based on alternative 2D nanomaterials?

Electrode fabrication: enhancing bare electrodes with nanomaterials

Nanomaterials are very appealing materials for development of innovative future biosensors. Exceptionally small size and high surface area to volume ratios permit intimate interaction with target biomolecules [34, 35]. By comparison, sensors uniquely based on bare GCE, PGE or SPCE show limitations. Carbon nanomaterials provide advantages for electrochemical carbon based nanosensors in pharmacology and biomedicine [36]. The use of fullerenes and carbon nanotubes in electrochemical biosensors [37] may be compromised by a tendency for zero- (0D) and one- (1D) dimensional nanostructures to self-aggregate. In contrast, two-dimensional materials have a higher surface to volume ratio offering better stability for molecular interactions [38]. Graphene, a 2D hexagonal carbon structure arranged from sp2 hybridized carbon atoms, is the fundamental carbon precursor for graphitic materials with other dimensionalities; folded into 0D fullerenes, rolled into 1D CNTs, or stacked into 3D graphite. Graphene’s unique properties include an atypically high specific surface area (2600 m2/g), mechanical strength (130 GPa) and electrical conductivity reaching 6000 S/cm in its pristine form. Various approaches have aimed to synthesize graphene, such as chemical vapor deposition (CVD), chemical exfoliation methods or physical vapor deposition (PVD) [39]. It has to be noticed that PVD is a relatively new field to grow 2D materials, and graphene produced by this technique exhibit excellent electrochemical properties [40]. Nowadays, the chemical oxidation of graphite followed by exfoliation in aqueous solvents to produce GO and/or its subsequent reduced form remains a popular pathway for generation of ultraperformance graphene electrochemical biosensors [41].

Nanomaterials used in biosensing vary in morphology (size, shape), mechanical, optical, electrical and chemical properties, biocompatibility and stability [42]. Diverse synthesis methods provide materials with various functionalizations and crystallinity, yet exactly how those parameters influence the electrochemical biosensing efficiency is rarely discussed. For elucidation of structure/reactivity relationships of electrochemical electrodes we take graphene as a benchmark.

Morphology, disorder and defects influence electrical conductivity of graphene: a critical aspect for biosensing

The variety of graphene, graphene composites and issues of nomenclature

In the broadest sense, graphenic nanomaterials encompass a variety of carbon sheets with different number of layers (single, few, multi) and a range of lateral sizes, oxygen content and functional groups, thus the term “graphene” warrants careful application. Recently, Wick et al. categorized different graphene types according to thickness, lateral dimension and atomic carbon/oxygen (C/O) ratio [43] (Fig. 1a), bringing more clarity in this field. Establishing standardization and nomenclature will be especially important to avoid misunderstanding between researchers, industry and authorities in the demanding context of healthcare applications. Key questions concern the influence of the layers, lateral size and C/O ratio on the electrochemical detection of molecules. Depending of the carbon source (e.g. graphite, CNT, SiC, agricultural waste) and synthesis method, various graphene forms with distinct qualities can be obtained, such as multilayer (ML) graphene [44, 45], graphene nanoribbons after unzipping the CNTs [46], graphene oxide, reduced graphene oxide (rGO) and graphene quantum dots (GQDs) [47]; plus functionalized graphene with various additional groups (COOH, NH2, etc.) [48, 49]. Scanning electron microscopy (SEM) reveals graphite and graphene nanoplatelets (Fig. 1b) as short piles of platelet-like graphene sheets, identical to those in the walls of CNTs, but in a planar structure [50]. GO and rGO modified GCE by drop-casting and electrochemical reduction, respectively, show different surface morphology (Fig. 1c). Reduced graphene oxide usually has an increased number of wrinkles which was found highly beneficial for the design of electrochemical biosensor with enhanced surface area. Introducing additional complexity, graphene can be mixed with nanoparticles or doped with various heteroatoms (e.g. N, B and S) [51], fabricated in 3D architectures or in a plethora of composites (e.g. foams), many applicable to electrochemical biosensing [52, 53].

Fig. 1
figure 1

Morphology of various graphene materials. A Classification grid for the categorization of different graphene types [43]. B TEM images of graphene nanoplates (1–5 nm), graphene nanoplates (5–20 nm), micro-graphite, and SEM image of natural graphite (top to bottom) [50]. C SEM images of GO (a) and ERGO (b) [118]

Engineered defects in graphene; understanding their role in electrochemistry

The electrochemical electrode response to reactants is mainly dependent on electron transfer kinetics and available surface area. The rate of heterogeneous electron transfer (HET), i.e. the transfer of electrons from/to electrode to/from molecules, is also correlated to the electrodes’s density of electronic states (DOS) and surface chemistry (number of defects, functional groups, impurities etc.). Introducing disorder/defects and roughness achieves a higher DOS, improving the HET of an electrode. It is established that the existence of edge plane sites/defects on graphitic materials improves the electron transfer processes [54] (Fig. 2a), occurring about 106 times faster at defects or edge-like sites versus basal or defect-free planes [55]. Faster electron dynamics at graphene edges result in excellent capacitive and electrocatalytic properties [56, 57] with four-fold specific capacitance and doubled current density found at the edge instead of the basal-plane. The HET rates of edge and basal planes of graphite are dependent on the type of inner and outer-sphere redox system used. For example, in the case of [Ru(NH3)6]3+/2+, the HET for both planes is similar, whereas for [Fe(CN)6]3−/4− it is dramatically different since the edge-plane activity is faster [55]. For the [Ru(NH3)6]3+/2+ redox system with an outer-sphere electron transfer mechanism, the charge transfer rate is mostly affected by the electronic properties of the electrode, in particular its DOS near the Fermi potential. In contrast, [Fe(CN)6]3−/4− represents an inner-sphere redox probe, thus its electron transfer kinetics is dependent on both surface microstructure and DOS.

Fig. 2
figure 2

(Reprinted with permission from [149]; copyright 2017, IOP Publishing)

Versatility of defects in graphene. a Schematic representation of a HOPG surface showing discrete basal plane and edge plane islands. Side on view of the HOPG surface highlighting its basal plane and edge plane like-sites/defects which exhibit contrasting behaviours in terms of electrochemical activity, where electron transfer kinetics of the latter are overwhelmingly dominant over that of the former which in comparison are relatively (electrochemically) inert. Reprinted from [54] with permission of The Royal Society of Chemistry. b An underlying graphene surface with few- and multi-layered graphitic islands, indicating the basal and edge plane electron transfer sites; notice the electrode surface is akin to a HOPG surface. Reprinted from [54] with permission of The Royal Society of Chemistry. Graphene atomic structure as obtained from density functional theory (DFT) calculation for: Stone–Wales defects (c), single vacancy defect (d), and multiple vacancy defects (e) in graphene. Reprinted with permission from [146]; copyright 2010 American Chemical Society. f Aberration-corrected annular dark-field scanning transmission electron microscopy (ADF-STEM) of line defects in graphene; scale bars: 5 Å. Reprinted with permission from [147]; copyright (2010) Nature Publishing Group. g Graphene in-plane heteroatom substitution defect model: nitrogen defects. Reprinted with permission from [148]; copyright (2010) American Chemical Society. h Illustrations of a graphene sample containing both point and line defects. The red regions define the structurally damaged area (S-region), and the green circles and lines are the activated area (A-regions) where the D band is active

Although single-layer pristine graphene of perfect 2D crystal structure has promising applications, its electrochemical activity suffers from an “Achilles’ heel” lack of an edge-plane band or significant surface defect densities [54]. Nonetheless, graphene acquires significant electrochemical properties when in few or multilayers structures (Fig. 2b). Existing defects (kinks, vacancies, steps) on the edge planes of epitaxial graphene (EG) generate localized edge states that result in high DOS near the Fermi level and increased electron transfer kinetics [58].

To monitor electrode modifications and, in particular, to access the presence of defects, Raman spectroscopy is a powerful, fast, non-destructive method, sensitive to both the electronic and phonon properties of the sample surface. Carbonaceous materials, characteristically generate a key quantifiable ratio between the spectrum’s D peak at approximately 1350 cm−1 and G peak at circa 1580 cm−1 [59]. The D peak corresponds to the breathing vibration of carbon in both sp3 bonds and sp2 rings on the edge-plane, while the G peak corresponds to the stretching vibration of carbon in both sp2 chains and rings in the basal-plane. The ratio of the D peak around 1350 cm−1 and G peak around 1580 cm−1 provides an estimate of the density of defects [60]. For the electrode pretreated by Prasad et al., the enhanced edge planes defect sites could also be confirmed by higher intensity ratios between D and G peaks in the Raman spectra [31]. For N layered graphene, the stacking order between layers has a significant influence on the band and interlayer phonon properties [60]. Meanwhile, a more disordered or defective structure induced frequency shifts and an increase in the linewidth of the Raman bands. Edges naturally exist in graphene whatever the technique used for synthesis, and they represent a kind of defect because the translational symmetry is broken. The lack of edge plane defects of pristine graphene can be highlighted by the absence of the D peak [43]. The high I(D)/I(G) ratio in GO confirms that its lattice is distorted and has a large amount of sp3-like defects caused by the oxidation process. The 2D peak between 2650 and 2700 cm−1 is, as the G peak, also characteristic of sp2 hybridized carbon–carbon bonds in graphene, and is extremely sensitive to defects, as well as, doping, thickness of graphene and nature of the substrate holder [60, 61]. The absence of 2D mode in GO primarily indicated a fully-disordered sp2 bonding structure mainly caused by functional groups. In contrast to mechanically exfoliated graphene, where the stacking order is well controlled, and 2D peak is a benchmark for the determination of the number of layers (NL), in the case of GO, the 2D peak is not related to NL. Few studies have been reported in this sense. Toh et al. has proposed an ingenious microinjection-micromanipulator system in order to estimate the heterogeneous electron transfer rate (ko) in function with the number of layers for IrCl 2−/3−6 probe [62]. On Si/SU8 substrate, the ko value augments with increasing NL: 3.08 × 10−3, 8.1 × 10−3, 1.06 × 10−2 cm s−1 for mono, few and multilayer graphene in the basal planes, respectively. On the edge/step between the few and multiplayer graphene flakes, the value was 9.88 × 10−2 cm s−1, showing a paramount importance to have edge plane defect sites in biosensing. As a reference, ko value was 1.93 × 10−2 cm s−1 for graphite. However, the substrate is believed to influence electrochemical properties and the k° value. As aforementioned, the synthesized graphene has a panel of morphology, and graphene tends to preferentially form aggregates to minimize the presence of edge plane sites. Smart approaches should be employed to avoid the morphological aggregation process [63]. Akhavan et al. has compared the DPV electrochemical activity of GCE, graphite, GO and rGO nanosheets, GO and rGO nanowalls (NWs). Due to the extremely sharp edges vertically aligned on GCE, the GO NWs and rGO NWs show an enhanced electrochemical reativity of the four free DNA bases, single-stranded (ss) DNA and double-stranded (ds) DNA [64].

Synthesized graphene and GO retain desirable defects [65, 66] (Fig. 2c–h) that can be divided into two categories; foreign adatoms and substitutional impurities. Defects can be introduced by different ways, such as ion-implantation, plasma treatment, functional groups. Substitutional impurities e.g. nitrogen, can form three chemical bonds and so take the place of carbon atoms in the graphene lattice. Graphene defects change the length of the interatomic valence bond and orbital, thereby altering its electrical properties. Point and single vacancy defects, unavoidable in current preparation methods, produce a decrease in the conductivity of graphene. Intrinsic defects aside, conductivity can also decrease due to the existence of oxygen groups and square resistance in GO can reach more than 1012 Ω. In contrast to oxygen, addition of different atoms such as nitrogen or boron, improved the conductivity of graphene [67]. In comparison to basal planes, the electrochemical activity of the graphene at its edge defective site is higher with an electron-transfer current from CV curves enhanced up to four orders of magnitude (Fig. 3a) [56]. Defects in GO can be either permanent or transient, the latter can be healed after a reductive restoration [66]. Self-healing graphene oxide/polymer composites are of great interest in the biomedical sensing field.

Fig. 3
figure 3

Structural properties of various graphene materials. a CVs of the basal plane (blue) and a graphene edge (red) in an aqueous solution of phosphate-buffered saline. Reprinted with permission from [57]. b XPS spectra of GO and rGO [135]. c XRD pattern of graphite (black), GO (red) and rGO (blue) [74]. d Raman spectra of GO (red) and rGO (blue) [74]. After reduction the intensity ratio between D and G (I(D)/I(G) increased

Comparing defect influence on graphene oxide versus reduced graphene oxide electrodes

GO is generally prepared by oxidation of graphite with subsequent exfoliation by sonication to obtain colloidal suspensions of mono-, bi-, or few-layer graphene oxide sheets [68]. Sonication effects on the mechanical properties and yield of GO is reported by various groups [69,70,71,72]. Helping control oxidation levels on the GO surface, Zeng et al. oxidized a graphite electrode using a one-step electrochemical approach by applying a positive potential between 0 and 3.0 V at 50 mVs−1 in 0.025 M phosphate buffer solution. Oxidation of graphite occurs at around 1.4 V and graphene oxide nanosheets could be obtained with good coverage for only 2–5 CV scans [73]. The surface of natural graphene is hydrophobic with little dispersibility in most solvents, whereas GO has hydroxyl and epoxide groups on the basal plane, plus carbonyl and carboxyl groups positioned at the sheet edges providing hydrophilicity for water dispersibility. By X-ray diffraction (XRD) was indicated that sonication increases the interlayer distance of graphene sheets confirming a good exfoliation effect [74]. Sehrawat et al. show using high-resolution-transmission electron microscopy (HRTEM) an obvious increase of the GO sheet transparency with an increase of ultrasonication time [75]. Subjected to sonication parameters of 40 kHz and 400 W, the size of the GO sheets may decrease sharply within the first hour of a sonication yet then remained relatively unchanged after 5 h of sonication, with a typical sheet surface area less than 200 µm2 [70]. Disaggregated into small flakes, GO sheets tend to restack together [70]. Dispersed GO is typically characterized by a 2:1 C/O ratio [76]. Unlike pristine graphene, defects on the GO surface greatly influence sensing performance by providing sites for strong interactions with charged species [77]. However, although GO surface groups help bind molecules avidly, it is less conductive, thus a reduction process is often performed to reach a compromise between reactive groups and conductivity.

Reduced GO can evolve from various reduction approaches including chemical, thermal, hydrothermal, microwave, microbial/bacterial, photo-chemical/photothermal, and electrochemical methods [78]. Reduction converts sp3 to sp2 carbon domains mediating important changes in the physico-chemical properties of GO (e.g. higher C/O atomic ratio, decrease of interlayer distance, Fig. 3b, c). The main reduction methods explored are chemical, thermal, hydrothermal and electrochemical [79]. For chemical reduction methods (CR), hydrazine hydrate (N2H4·H2O) is the most used agent, although other reductants have been employed with good efficiency [68]. Thermal reduction (TR) is generally performed by heating GO above 1000 °C [68]. However, some recent studies reported good reduction by annealing using temperature below 350 °C [80, 81]. Hydrothermal reduction (HR) is a green method in comparison to CR and TR because the use of moderate temperature and pressure in aqueous solution [82]. Amongst the numerous reduction methods, the electrochemical approach is a fast, easy, economic, scalable, and environmentally friendly route [62].

Typically, reduction of GO is accompanied by smaller sp3 domain sizes compensated by more sp2 domains, indicated by a higher ratio between the intensity of D and G peaks and a narrower D band in rGO compared to GO in the Raman spectra (Fig. 3d). Carbon material versatility reflects strong interdependence between physical properties and the ratio of sp2 (graphite like) to sp3 (diamond like) bonds. A three stage amorphization trajectory model has been proposed to describe the introduction of a series of defects in a perfect graphene sheet, transitioning from: (1) graphite to nanocrystalline graphite (nc-G); (2) nc-G to amorphous carbon (a-C); a-C to tetrahedral amporphous carbon (ta-C) [83]. The intensity ratio I(D)/I(G), an indicator of carbon network order or disorder, has been used to estimate lateral crystallite size La according to a Tuinstra-Koening relation. Although this ratio reportedly varies in an inversely proportional manner to La [84], this did not strictly concord with the observations of Ferrari et al., since one could expect presence of sp3 as well as sp2 domains during progressive reduction of graphene [83, 85]. Rather, the evolution of I(D)/I(G) with the crystallite size was in perfect agreement with predictions for the stage 2 transition from nc-G to a-C [83], while the Tuinstra-Koening relation was only valid for the stage 1 graphite to nc-G trajectory. Reduction of GO may provide abundant structural defects and improve drastically the electrochemical activity of bare and GO modified electrodes. However, contrary to convention, some have reported that GO had a better CV signal, even though GO itself should be less conductive [86].

The improved electrical conductivity from partial restoration of π-conjugated sp2 structure, with more chemically reactive defective sites make rGO a favoured graphene material for the detection of DNA. Direct comparison of GO vs rGO for DNA adsorption measured by fluorescence spectroscopy revealed a better sensing ability for rGO due to a richer sp2 carbon surface allowing better π–π stacking with DNA bases [87]. Scanning electron microscopy images of electrochemically reduced GO reveal an irregularly crumbled and sheet-like structure with an increased density and thickness proportional to the number of cycles [88]. Consequently, this increased the effective surface area and improved the conductivity. Based on the Randles–Sevcik equation, GO has a higher active surface area with a better electrochemical reaction ability [89,90,91]. It is possible to reversibly control the introduction of defects at graphene boundaries [92]. Monitoring the oxygen content and fraction ratio of sp2 to sp3-hybridized carbon atoms can be used to tune the carrier mobility by over 12 orders of magnitude and though residual oxygen in rGO prevents carrier mobilities from equating to pristine graphene values, high electron mobilities over 1000 cm2 V−1 s−1 have been recorded in thin rGO films [93]. However, rGO seems chemically less stable than graphene and GO [94,95,96]. It is therefore clear from recent studies that each graphene material type has advantages and drawbacks with debate as to which one could prove best for biosensing.

GO and rGO could be assembled onto the electrode by various techniques such as drop-casting, dip-coating, spray coating and layer-by-layer deposition [1]. A film forms through π–π electronic interaction between graphene and the bare electrode [97]. The electrochemical reduction (ER) of GO could be realized in various electrolytes. Two different pathways are generally undertaken; the one-step path involves the direct electrochemical reduction of graphene oxide in colloidal solution by a supporting electrolyte on the electrode. The two-step method involves the pre-deposition of GO onto the electrode before its reduction using a conductive electrolyte on the electrode. Table 1 shows recent protocols used for the preparation of graphene modified GCE, PGE and SPCE.

Table 1 Platforms using rGO or GO on glassy carbon electrode, pencil graphite electrode and screen printed carbon electrode

Sensors comparing oxidized graphene and reduced graphene oxide

Applying a CV potential of up to 2.0 V for 500 s to oxygenated epitaxial graphene in PBS (anodization process), a mixture of DNA bases could be detected simultaneously, without a pre-hydrolysis step. dsDNA was discriminated from ssDNA with better sensitivity than when using GCE and boron doped diamond. Moreover, mixtures of biomolecules (AA, DA and UA) were treated as individual peaks [58]. This was associated with both an increase in the electron transfer rate (0.0101 cm/s) and edge plane defects. The latter were correlated to both an increase in D peak intensity (higher I(D)/I(G) ratio) and oxygen content. The results were supported by both qualitative (using both [Fe(CN)6]3−/4− and [Ru(NH3)6]3+/2+ redox probes) and quantitative analysis (reversible process through Randles–Sevcik and Nicholson equations). An ability to controllably change the anodization time, served to accurately tune the graphene defect density. It was clearly shown that anodized EG at 200 s had a slower HET rate constant than the one anodized at 500 s. The latter had more defective sites, a sharper redox [Fe(CN)6]3−/4− peak separation and, higher capacitance in potassium chloride (KCl) at 0.025 V.

Zhou et al. compared the electrochemical sensing ability of CR-GO/GCE (chemically reduced GO) with graphite/GCE and GCE with a different morphology (Fig. 4a) [106]. Their DPV measurement indicated that CR-GO/GCE had a better sensing performance to detect DNA nucleotides, ss and ds. Moreover, single-base mismatches (G→A or C→T mutations) within single-nucleotide polymorphisms (SNPs) and multiplexed DNA nucleotides could be accurately detected using CR-GO/GCE. Other molecules such as glucose or β-nicotinamide adenine dinucleotide (NADH) were also detected by such methods. The enhanced electrochemical properties could be attributed to its single-sheet nature, high conductivity, a better reaction ability (apparent electrode area A ~ 0.092 cm2), lower charge transfer resistance (160.8 Ω), antifouling properties and a higher density of edge-plane defect sites (I(D)/I(G) ~ 1.38). Edge-plane defects provided numerous active sites valuable for accelerating electron transfer between the electrode and species in solution. However, it was clear from an I(D)/I(G) of graphite/GCE close to 0 [27], that defects were not the only contributing factor for fast electron transfer. Unique characteristics of CR-GO/GCE presumably made the electron transfer easier.

Fig. 4
figure 4

Comparison of bare electrode, graphite electrode and various graphene electrodes on electrochemical sensing. A (a) DPVs at the GC electrode for G (blue), A (orange), T (violet), and C (magenta), respectively; (b) DPVs at the graphite/GC electrode for G (blue), A (orange), T (violet), and C (magenta), respectively; (c) DPVs at the CR-GO/GC electrode for G (blue), A (orange), T (violet), and C (magenta), respectively; (d) DPVs for a mixture of G, A, T, and C at CR-GO/GC (green), graphite/GC (red), and GC electrodes (black); (e) DPVs for ssDNA at CR-GO/GC (green), graphite/GC (red), and GC electrodes (black); (f) DPVs for dsDNA at CR-GO/GC (green), graphite/GC (red), and GC electrodes (black). Concentrations for different species (a-f): G, A, T, C, ssDNA or dsDNA: 10 μg/mL. Electrolyte: 0.1 M pH 7.0 PBS. Reprinted with permission from [106]; Copyright 2009 American Chemical Society. B Histogram representing a comparison of impedimetric signals on GPO, GO, TR-GO, ER-GO and bare DEP electrode. Signal is represented as average Rct ratio ((Rct protein − Rct blank)/(Rct aptamer − Rct blank)). Error bars correspond to triplicate experiments. All measurements were performed with 10 mM K4[Fe(CN)6]/K3[Fe(CN)6] in Tris buffer solution (pH = 8.2) at room temperature. Reprinted from [107] with permission of The Royal Society of Chemistry. C The comparison of impedimetric response on graphene oxide, ER-GO and TR-GO recorded after hybridization with wild-type (grey), mutant (purple) and non-complementary (black) target DNA sequences. The signal is represented as Δratio = Δt/Δp, Δt = Rct, target-Rct, blank, Δp = Rct, probe-Rct, blank. Standard deviations correspond to triplicate experiments. Reprinted from [108] with permission of The Royal Society of Chemistry. D (a) Raman spectra of ERGO film at different CV cycles; (b) CV curves containing 30 µM isoniazid (INZ) at different reduction cycles, and (c) line chart of the relationship between reduction cycle number and peak current response towards INZ [118]

One of the most informative direct comparisons was undertaken by Báez et al., who compared the electrochemical performance of graphene oxide reduced by chemical (CRGO), thermal (TRGO), hydrothermal (HRGO), and electrochemical (ERGO) methods taken from a similar GO source [79]. The reduced GO samples had obvious morphological differences, with specific surface areas, oxygen contents, and electrochemical activity (Fig. 4b). TRGO had the best sensitivity for the detection of guanine by DPV measurements. The performance of TRGO likely reflected its dramatically different morphology (highest porosity and surface area), highest number of electroactive sites (more significant CV current), and good electrical conductivity (one of the highest C/O ratio). However, in contrast CR-GO/GCE, it was unlikely to be related to common defects activated in the Raman spectra, since TRGO had the lowest I(D)/I(G) ratio. Also, although their electron transfer could be fast, it was not the fastest with regard to the redox peak separation. In contrast, by comparing graphite oxide, GO, TRGO and ERGO dried-absorbed on DEP-chips (disposable electrical printed chip), GO provided the best sensitivity for the detection of the important blood coagulation protein thrombin [107] (Fig. 4c). From similar experimental comparisons, in the same research group, GO on DEP-chips was the best candidate for the detection of SNPs correlated to Alzheimer’s disease [108] (Fig. 4d). Despite convincing experimental evidence using only impedimetric measurements, no surface and structural analysis was provided. Giovanni et al. gave a possible explanation of a better sensitivity for GO (e.g. presence of defects and oxygen groups) claiming agreement with the earlier studies of Lim et al. [58]. However, a comparison of XPS spectra clearly indicated that the C/O ratio of GO for each study was drastically different. Examples of such differences between laboratories using broader electrochemical methods for detection and with greater emphasis on the quantitative analysis (spectroscopy, calculation of charge transfer resistance, etc.) could bring more clarity to this field.

Functionalization and composite materials broadening the scope of electrochemical biosensor tunability

Structurally modifying graphene through chemical and physical functionalization methods revealed the numerous possibilities for tuning its structure [109]. By performing EIS measurement, Bonnani et al. showed that chemically reduced GO with perpendicularly grafted carboxyl groups, mainly at edge sites (CRGO-COOH), dramatically outperformed the DNA sensing performance of GO as a result of a more efficient immobilization of the probes [48]. The probe density estimated from chronocoulometry was 2.41 × 1013 and 3.76 × 1013 molecules cm−2 for GO and CRGO-COOH, respectively. Interestingly, COOH groups on CRGO did not change its electrochemical response significantly (CV and EIS curves) meaning no significant damage was introduced during grafting. This seemed in good agreement with the fact that the grafted COOH groups did not introduced many defects in the structure as indicated by Raman measurements [110, 111].

The concept of combining graphene with other materials to enhance its electrochemical properties is very compelling, taking advantage of synergistic effects for enhanced biosensors. Several methods have been developed for the preparation of less aggregated and more stable graphene electrochemical biosensors from graphene-based composites. Less graphene aggregation not only provides a high active site density but also increases the surface area and porosity, both extremely beneficial for the anchorage of biological molecules. For example, graphene has been found to be an excellent 2D surface to incorporate Au, Pt and Pd nanoparticles for applications in energy, catalysis and sensing, by way of providing improved charge transfer resistance and peak current intensities. Nonetheless, many challenges remain to be resolved [112]. Nanoparticles are often electrodeposited on the surface of graphene. For example, PtAu nanoclusters could be deposited on rGO in aqueous solution containing 0.2 M H2SO4, 0.5 mM HAuCl4 and 0.5 mM H2PtCl6 with a deposition time and potential of 400 s and − 0.2 V, respectively [97]. Gold nanoparticles (AuNPs) modified graphene electrodes are probably the most widely used for sensing devices due to their unique optical and electronic properties, stability, low cytotoxicity and relative ease of synthesis with a typical size ranging from 3 to 200 nm. This hybrid material could be used in almost any domain of sensing, ranging from optical fluorescence to electrochemical approaches. For example, by modifying PGE with AuNPs, Mandli et al. designed a highly sensitive and selective electrochemical microRNA (miRNA) biosensor with a LOD of 100 pM and a linear range from 200 to 388 nM [113].

Nowadays, complex graphene composites biosensors are elaborated but quantitative analyses correlating how charge electron transfer affects the biosensors have yet to be completed. In summary, a good electrode needs to balance conductivity and defect density [114]. Also, it is clear that the surface area, morphology, chemistry, functionalization and newly developed hybrid graphene materials play an important role towards high efficient biosensors. Table 2 highlights studies concerning the structural, chemistry and electron transfer properties of electrochemical biosensors based on graphene and graphene composites.

Table 2 Influence of morphology, defects and conductivity on electron transfer properties of electrochemical biosensors based on graphene and graphene composites

Aspects affecting stability and/or reproducibility of electrochemical biosensors

Graphene’s discovery in 2004 introduced unique properties revolutionizing preconceptions that such 2D materials were thermodynamically unstable. However, reliable stability and biosensor reproducibility remain a major problem for consistent performance within large-scale technological applications. Electrochemical sensor device reliability is influenced by limitations concerning electrode pretreatments, precise synthesis of graphenic materials [119] and device architectures. The sensing performance may differ from device-to-device even though the graphene materials originate from the same batch and fabrication protocol. Key factors that could influence the stability and reproducibility, taking graphene as a benchmark material are portrayed in Fig. 5a.

Fig. 5
figure 5

(Reprinted with permission from [86]; copyright 2017 Elsevier)

Reproducibility and stability of graphene electrodes for biosensing. a Factors that influence the stability and reproducibility of graphene biosensors. b CV curves of reduced graphene oxide by two-step approach (i.e. reduction after drop-casting) after different acquisition time obtained in 0.1 M KCl solution containing 1 mM K3[Fe(CN)6]/K4[Fe(CN)6]. c The representative scheme of experimental procedure followed in the hybridization occured between miRNA-34a target and its complementary DNA probe at the surface of CA/GO/PGEs

Conversely, GO is a metastable material that may undergo spontaneous reduction under certain conditions such as exposure to sunlight [11] and its stability and aggregation have been investigated in various media. In aqueous solution, ionic strength, pH, ion valence and presence of organic matter all significantly influence GO properties in solution. Its stability is also strongly influenced by the interdependent factors of surface oxidation and polarity. Regarding polarity, a GO sheet is amphiphilic, bearing hydrophilic sites (–COOH at the edge plus phenol and hydroxyl groups at the basal plane) and graphitic domain hydrophobic sites at the basal plane. Furthermore, the colloidal stability of GO sheets depends on their edge-to-edge interacting geometry. Electrostatic repulsion tends to dominate, but if brought face-to-face, the attractive Van der Waals and π–conjugated interactions may predominate [120]. This helps explain the aggregative behavior of GO during storage and simple GO redispersion by sonication. Disaggregated into small flakes, GO sheets tend to restack together [70]. The stability and reproducibility of graphene electrodes, are strongly influenced by the concentration of GO sheets and extent of dispersion in solution. Good dispersion is thus already a critical parameter to take into account, as are the polarity of GO sheets and pH of the solution. Cote et al. showed by Zeta potential measurements that the hydrophilicity of GO increased with pH and decreasing the sheet size from a micron to a nanometer scale (< 100 nm) lowering the extent of reduction [120].

The stability of graphene oxide in water is affected by subsequent reduction of its functional groups, and by pH which is known to have the greatest influence on rGO stability [11]. Figure 5b illustrates a significative change of rGO CV signal over time. Nafion and lysozyme are good dispersants for rGO [121, 122]. Functionalization could increase the performance of carbon-based materials. Graphene based composites with polymers or inorganic compounds may offer more sensitivity, selectivity and reproducibility. The combination of 2D materials with appropriate interface materials represents a successful approach to improve the sensitivity and stability of a biosensing platform [123]. Important binding agent members such as chitosan, dopamine or nafion can increase graphenic stability and fix molecules tightly onto the electrode surface [2, 5, 44], thereby enhancing the adherence of GO onto the electrode surface. Composites made by mixing graphene with metallic nanoparticles may open new doors towards highly stable and reproducible biosensors. In a complementary manner, chemical activation procedures may help stabilize the surfaces of electrodes and provide better attachment for graphene onto surface [124] (Fig. 5c). When not used, it is generally advised that graphene electrodes be kept at 4 °C for stability [106], as recommended for other 2D materials [18]. Electrochemical measurements based on field-effect (FET) technology encouraged interest in 2D materials for sensors due to their sensitivity, fast response and real-time monitoring. New composite formulations can resolve problems such as the fragility of black phosphorus, readily degraded by water vapor, oxygen and irradiation from visible light [123]. By doping BP with sulphur, the innovative FET device still retained useful conductivity after exposure to air for 21 days with a decrease of charge-carrier mobility of only 22.6% [125].

With respect to device architecture, clear electrochemical differences are introduced by the choice of aptamer binding method, either by simple adsorption (Table 3) or involving covalent interaction (Table 4). Regarding DNA probe immobilization at the surface of the WE by simple adsorption, hexagonal cells of graphene sheets and the aromatic rings of nucleobases share Van der Waals forces, ionic interactions, π–π stacking, and hydrogen bonds. DNA is physically adsorbed better on reduced GO than GO, since rGO has more aromatic regions for π–π stacking interactions with DNA bases and a lower surface charge density lessens electrostatic repulsion of the negatively charge DNA backbone [87, 107, 126].

Table 3 Biosensors with nucleic acid aptamer immobilization using noncovalent approach
Table 4 Biosensors with nucleic acid aptamer immobilization covalent approach

Notably, when using [Fe(CN)6]3−/4− as the redox probe, an increased impedimetric response upon physical adsorption of the DNA probe preceded a decreased impedance signal after hybridization with the target. Inferred partial release of the dsDNA from the electrode could decrease the total charge present on the electrode surface, thus reducing the charge transfer resistance [127]. Nonetheless simple adsorption techniques can be advantageously mild and Raman analysis indicated no introduced defects or disorder on the graphene structure [128].

In contrast, covalent bonding immobilizes the DNA probes at one end, improving stability with protection from desorption. This approach can achieve good vertical orientation of the biomolecules on the electrode surface, favouring efficient aptamer/analyte hybridization. Unlike adsorption-based methods, the charge transfer resistance continued to increase upon analyte hybridization due to a more negatively charged species on the surface of the electrode and repulsion of [Fe(CN)6]3−/4− by the negatively charged phosphate backbone of the hybrdized DNA probe [129]. This improved sensitivity for electrochemical analyte detection.

Conclusions

Ubiquitous development of electrochemical biosensors for healthcare and biomedicine has yet to produce fully integrated functional devices approved for real clinical evaluation. Nonetheless, the growing portfolio of biosensors using unique versatile 2D nanomaterials and their hybrid composites as a thin surface veil imparting transformative properties has encouraging potential for the biomedical field. Manufacturing graphene sheets to perfection need not be essential, so long as measurement conditions and quality control can maintain a satisfactory consistency and reproducibility. Although defects are often considered detrimental to material properties, if engineered to tailor the surface in a controlled manner, they can introduce properties that enhance function for a wide range of new electrochemical biosensors.

Research into innovative 2D material-based sensors beyond the first application of graphene in 2008, has provided useful experimental examples, yet for commercialization the technology is still in its infancy, facing challenging fabrication procedures and poor control of uniformity and reproducibility for reaching the desired properties [130]. There is still need for much research into fundamental aspects of this field. Reports of biosensors demonstrating relatively good stability and reproducibility often combine graphene with composite compounds [131], modulate graphene via doping [132], introduce covalent approaches for target detection and/or place emphasis on carefully controlled storage conditions. It is anticipated that progress in understanding the mechanism for the excellent electro-catalytic activity of 2D materials will improve prospects for meeting the urgent need for point of care (POC) devices [133] and live cell monitoring [134], through low-cost miniaturized potentiostats. Biosensor nanotechnology provides one of the most pertinent examples where success depends upon multidisciplinary collaboration supported by partnership between public research centres, industry and the health sector.

Availability of data and materials

Not applicable.

Abbreviations

0D:

zero-dimensional

1D:

one-dimensional

2D:

two-dimensional

3D:

three-dimensional

AA:

ascorbic acid

ABS:

acetate buffer solution

AC:

alternating current

a-C:

amorphous carbon

aM:

attomolar

AMEL:

amelogenin

AMGX:

proprietary name for AMEL complementary probe sequence

APBA:

3-aminophenylboronic acid

Apt:

aptamer

APTES:

3-aminopropyltriethoxysilane

AuE:

gold-wire electrode

AuNPs:

gold nanoparticles

BP:

black phosphorus

CA:

chronoamperometry

CB:

carbon black

CCG:

chemical converted graphene

cDNA:

complementary DNA

CE:

counter electrode

cm:

centimeter

CNTs:

carbon nanotubes

CR:

chemical reduction

CRGO:

chemically reduced graphene oxide

CV:

cyclic voltammetry

CVD:

chemical vapor deposition

DA:

dopamine

DC:

direct current

DEP-chip:

disposable electrical printed chip

DNA:

deoxyribonucleic acid

DOS:

density of state

DPV:

differential pulse voltammetry

dsDNA:

double-stranded DNA

E:

electrode potential

EDC:

1-ethyl-3(3-dimethylaminopropyl)-carbodiimide

EG:

epitaxial graphene

EIS:

electrochemical impedance spectroscopy

ER:

electrochemical reduction

ERGO:

electrochemically reduced graphene oxide

ET:

estriol

[Fe(CN)6]3−/4− :

ferrocyanide/ferricyanide ions

FET:

field-effect transistor

fM:

femtomolar

GCE:

glassy carbon electrode

GO:

graphene oxide

GOx:

glucose oxidase

GQDs:

graphene quantum dots

Gr-HPHT:

graphene-high pressure, high temperature

G-SL:

single-layer graphene

HBV:

hepatitis B virus

HET:

heterogeneous electron transfer

HIV 1:

human immunodeficiency virus 1

hpDNA:

hairpin DNA

HRGO:

hydrothermally reduced graphene oxide

Hz:

Hertz

I:

Current

ID/IG ratio:

ratio of the intensity of D and G Raman peaks

K3[Fe(CN)6]:

potassium ferricyanide

K3PO4 :

tripotassium phosphate

K4[Fe(CN)6]:

potassium ferrocyanide

KCl:

potassium chloride

ko :

heterogenous electron transfer rate constant

Ks:

electrode transfer rate

La :

lateral crystallite size

LOD:

limit of detection

M:

molar

MES:

2-(N-morpholino)ethanesulfonic acid

mg:

milligram

MgCl2 :

magnesium chloride

miRNA:

microRNA

mL:

milliliter

ML:

multilayer

mM:

millimolar

MnTPP:

manganese (III) tetraphenylporphyrin

MoS2 :

molybdenum disulfide

MRSA:

methicillin-resistant Staphylococcus aureus

mV:

millivolts

mV/s:

millivolts per second

MWCNTs:

multi-wall carbon nanotubes

N/A:

not applicable/available

N2 :

nitrogen

Na2SO4 :

sodium sulfate

Na3PO4 :

trisodium phosphate

NaCl:

sodium chloride

NADH:

nicotinamide adenine dinucleotide

NaOH:

sodium hydroxide

nc-G:

nanocrystalline graphite

NHS:

N-hydroxysulfosuccinimide

NL:

number of layers

nm:

nanometer

nM:

nanomolar

PBS:

phosphate buffer solution

PCR:

polymerase chain reaction

pDNA:

probe DNA

pg:

picogram

PGE:

pencil graphite electrode

pM:

picomolar

POC:

point of care

PTCA:

3,4,9,10-perylene tetra-carboxylic acid

PVD:

physical vapor deposition

QDs:

quantum dots

Rct:

charge transfer resistance

RE:

reference electrode

rGO:

reduced graphene oxide

RNA:

ribonucleic acid

RSD:

relative standard deviation

RT:

room temperature

[Ru(NH3)6]3+/2+ :

hexaammineruthenium (III)/(II)

s:

second

S/N:

signal-to-noise ratio

SBS:

salbutamol sulfate

SiC:

silicon carbide

SNPs:

single-nucleotide polymorphisms

SPCE:

screen printed carbon electrode

SPE:

screen printed electrode

SSC:

saline sodium citrate

ssDNA:

single-stranded DNA

SWV:

square wave voltammetry

t:

time

ta-C:

tetrahedral amorphous carbon

TBS:

tris–HCl buffered saline

Thi:

thionine

THR:

thrombin

TMDs:

transition metal dichalcogenides

TRGO:

thermally reduced graphene oxide

TSC:

trisodium citrate

UA:

uric acid

V:

volts

VS2 :

vanadium disulfide

WE:

working electrode

zM:

zeptomolar

µA:

microampere

µF:

microfarad

µg:

microgram

µL:

microliter

µM:

micromolar

ϕ:

phase

ω:

frequency

Ω:

ohm

References

  1. Terse-Thakoor T, Badhulika S, Mulchandani A. Graphene based biosensors for healthcare. J Mater Res. 2017;32(15):2905–29.

    Article  CAS  Google Scholar 

  2. Maduraiveeran G, Sasidharan M, Ganesan V. Electrochemical sensor and biosensor platforms based on advanced nanomaterials for biological and biomedical applications. Biosens Bioelectron. 2018;103:113–29.

    Article  CAS  PubMed  Google Scholar 

  3. Suvarnaphaet P, Pechprasarn S. Graphene-based materials for biosensors: a review. Sensors. 2017;17(10):2161.

    Article  CAS  PubMed Central  Google Scholar 

  4. Gao N, Gao T, Yang X, Dai X, Zhou W, Zhang A, et al. Specific detection of biomolecules in physiological solutions using graphene transistor biosensors. Proc Natl Acad Sci. 2016;113(51):14633–8.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  5. Huang Y, Xu J, Liu J, Wang X, Chen B. Disease-related detection with electrochemical biosensors: a review. Sensors (Basel). 2017;17:10.

    Google Scholar 

  6. Bollella P, Fusco G, Tortolini C, Sanzò G, Favero G, Gorton L, et al. Beyond graphene: electrochemical sensors and biosensors for biomarkers detection. Biosens Bioelectron. 2017;89:152–66.

    Article  CAS  PubMed  Google Scholar 

  7. Shavanova K, Bakakina Y, Burkova I, Shtepliuk I, Viter R, Ubelis A, et al. Application of 2D non-graphene materials and 2D oxide nanostructures for biosensing technology. Sensors. 2016;16(2):223.

    Article  PubMed  CAS  PubMed Central  Google Scholar 

  8. Hu Y, Huang Y, Tan C, Zhang X, Lu Q, Sindoro M, et al. Two-dimensional transition metal dichalcogenide nanomaterials for biosensing applications. Mater Chem Front. 2017;1(1):24–36.

    Article  CAS  Google Scholar 

  9. Vlăsceanu GM, Amărandi R-M, Ioniță M, Tite T, Iovu H, Pilan L, et al. Versatile graphene biosensors for enhancing human cell therapy. Biosens Bioelectron. 2018;117:283–302.

    Article  PubMed  CAS  Google Scholar 

  10. Bhimanapati GR, Lin Z, Meunier V, Jung Y, Cha J, Das S, et al. Recent advances in two-dimensional materials beyond graphene. ACS Nano. 2015;9(12):11509–39.

    Article  CAS  PubMed  Google Scholar 

  11. Grieshaber D, MacKenzie R, Voeroes J, Reimhult E. Electrochemical biosensors-sensor principles and architectures. Sensors. 2008;8(3):1400–58.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Elgrishi N, Rountree KJ, McCarthy BD, Rountree ES, Eisenhart TT, Dempsey JL. A practical beginner’s guide to cyclic voltammetry. J Chem Educ. 2017;95(2):197–206.

    Article  CAS  Google Scholar 

  13. Rackus DG, Shamsi MH, Wheeler AR. Electrochemistry, biosensors and microfluidics: a convergence of fields. Chem Soc Rev. 2015;44(15):5320–40.

    Article  CAS  PubMed  Google Scholar 

  14. Hughes G, Pemberton RM, Nicholas P, Hart JP. Fabrication of miniaturised screen-printed glucose biosensors, using a water-based ink, and the evaluation of their electrochemical behaviour. Electroanalysis. 2018;30(8):1616–20.

    Article  CAS  Google Scholar 

  15. McCreery RL. Advanced carbon electrode materials for molecular electrochemistry. Chem Rev. 2008;108(7):2646–87.

    Article  CAS  PubMed  Google Scholar 

  16. Yang J, Gunasekaran S. Electrochemically reduced graphene oxide sheets for use in high performance supercapacitors. Carbon. 2013;51:36–44.

    Article  CAS  Google Scholar 

  17. Randviir EP, Brownson DAC, Metters JP, Kadara RO, Banks CE. The fabrication, characterisation and electrochemical investigation of screen-printed graphene electrodes. Phys Chem Chem Phys. 2014;16(10):4598–611.

    Article  CAS  PubMed  Google Scholar 

  18. Anderson K, Poulter B, Dudgeon J, Li S-E, Ma X. A highly sensitive nonenzymatic glucose biosensor based on the regulatory effect of glucose on electrochemical behaviors of colloidal silver nanoparticles on MoS2. Sensors. 2017;17(8):1807.

    Article  CAS  PubMed Central  Google Scholar 

  19. Cinti S, Arduini F. Graphene-based screen-printed electrochemical (bio) sensors and their applications: efforts and criticisms. Biosens Bioelectron. 2017;89:107–22.

    Article  CAS  PubMed  Google Scholar 

  20. Manjunatha JG. Electroanalysis of estriol hormone using electrochemical sensor. Sens Bio-Sens Res. 2017;16:79–84.

    Article  Google Scholar 

  21. Amare M, Menkir G. Differential pulse voltammetric determination of salbutamol sulfate in syrup pharmaceutical formulation using poly (4-amino-3-hydroxynaphthalene sulfonic acid) modified glassy carbon electrode. Heliyon. 2017;3(10):e00417.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Purushothama HT, Nayaka YA, Vinay MM, Manjunatha P, Yathisha RO, Basavarajappa KV. Pencil graphite electrode as an electrochemical sensor for the voltammetric determination of chlorpromazine. J Sci. 2018;3(2):161–6.

    Google Scholar 

  23. Cesiulis H, Tsyntsaru N, Ramanavicius A, Ragoisha G. The Study of Thin Films by Electrochemical Impedance Spectroscopy. In: Tiginyanu I, Topala P, Ursaki V, editors. Nanostructures and thin films for multifunctional applications: technology, properties and devices. Cham: Springer; 2016. p. 3–42.

    Chapter  Google Scholar 

  24. Braiek M, Rokbani KB, Chrouda A, Mrabet B, Bakhrouf A, Maaref A, et al. An electrochemical immunosensor for detection of staphylococcus aureus bacteria based on immobilization of antibodies on self-assembled monolayers-functionalized gold electrode. Biosensors. 2012;2(4):417–26.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  25. Ortiz-Aguayo D, Del Valle M. Label-free aptasensor for lysozyme detection using electrochemical impedance spectroscopy. Sensors (Basel). 2018;18:2.

    Google Scholar 

  26. David IG, Popa DE, Buleandra M. Pencil graphite electrodes: a versatile tool in electroanalysis. J Anal Methods Chem. 2017;2017:1905968.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  27. Yi Y, Weinberg G, Prenzel M, Greiner M, Heumann S, Becker S, et al. Electrochemical corrosion of a glassy carbon electrode. Catal Today. 2017;295:32–40.

    Article  CAS  Google Scholar 

  28. Yamanaka K, Vestergaard MDC, Tamiya E. Printable electrochemical biosensors: a focus on screen-printed electrodes and their application. Sensors. 2016;16(10):1761.

    Article  CAS  PubMed Central  Google Scholar 

  29. Navratil R, Kotzianova A, Halouzka V, Opletal T, Triskova I, Trnkova L, et al. Polymer lead pencil graphite as electrode material: voltammetric, XPS and Raman study. J Electroanal Chem. 2016;783:152–60.

    Article  CAS  Google Scholar 

  30. Gowda JI, Hurakadli GS, Nandibewoor ST. Pretreated graphite pencil electrode based voltammetric sensing of albendazole. Anal Chem Lett. 2017;7(3):389–401.

    Article  CAS  Google Scholar 

  31. Sudhakara Prasad K, Muthuraman G, Zen J-M. The role of oxygen functionalities and edge plane sites on screen-printed carbon electrodes for simultaneous determination of dopamine, uric acid and ascorbic acid. Electrochem Commun. 2008;10(4):559–63.

    Article  CAS  Google Scholar 

  32. Flexer V, Marque M, Donose BC, Virdis B, Keller J. Plasma treatment of electrodes significantly enhances the development of anodic electrochemically active biofilms. Electrochim Acta. 2013;108:566–74.

    Article  CAS  Google Scholar 

  33. Yammouri G, Mandli J, Mohammadi H, Amine A. Development of an electrochemical label-free biosensor for microRNA-125a detection using pencil graphite electrode modified with different carbon nanomaterials. J Electroanal Chem. 2017;806:75–81.

    Article  CAS  Google Scholar 

  34. Yakoh A, Pinyorospathum C, Siangproh W, Chailapakul O. Biomedical probes based on inorganic nanoparticles for electrochemical and optical spectroscopy applications. Sensors. 2015;15(9):21427–77.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  35. Su H, Li S, Jin Y, Xian Z, Yang D, Zhou W, et al. Nanomaterial-based biosensors for biological detections. Arch J. 2017;3:19–29.

    Google Scholar 

  36. Kurbanoglu S, Ozkan SA. Electrochemical carbon based nanosensors: a promising tool in pharmaceutical and biomedical analysis. J Pharm Biomed Anal. 2018;147:439–57.

    Article  CAS  PubMed  Google Scholar 

  37. Sherigara BS, Kutner W, D’Souza F. Electrocatalytic properties and sensor applications of fullerenes and carbon nanotubes. Electroanalysis. 2003;15(9):753–72.

    Article  CAS  Google Scholar 

  38. Hou S, Zhang A, Su M. Nanomaterials for biosensing applications. Basel: Multidisciplinary Digital Publishing Institute; 2016.

    Book  Google Scholar 

  39. Tite T, Donnet C, Loir A-S, Reynaud S, Michalon J-Y, Vocanson F, et al. Graphene-based textured surface by pulsed laser deposition as a robust platform for surface enhanced Raman scattering applications. Appl Phys Lett. 2014;104(4):041912.

    Article  CAS  Google Scholar 

  40. Fortgang P, Tite T, Barnier V, Zehani N, Maddi C, Lagarde F, et al. Robust electrografting on self-organized 3D graphene electrodes. ACS Appl Mater Interfaces. 2016;8(2):1424–33.

    Article  CAS  PubMed  Google Scholar 

  41. Szunerits S, Boukherroub R. Graphene-based biosensors. Interface focus. 2018;8(3):20160132.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Chimene D, Alge DL, Gaharwar AK. Two-dimensional nanomaterials for biomedical applications: emerging trends and future prospects. Adv Mater. 2015;27(45):7261–84.

    Article  CAS  PubMed  Google Scholar 

  43. Wick P, Louw-Gaume AE, Kucki M, Krug HF, Kostarelos K, Fadeel B, et al. Classification framework for graphene-based materials. Angew Chem Int Ed. 2014;53(30):7714–8.

    Article  CAS  Google Scholar 

  44. Teixeira S, Ferreira NS, Conlan RS, Guy OJ, Sales MGF. Chitosan/AuNPs modified graphene electrochemical sensor for label-free human chorionic gonadotropin detection. Electroanalysis. 2014;26(12):2591–8.

    Article  CAS  Google Scholar 

  45. Filik H, AslhanAvan A, Aydar S. Electrochemical determination of nicotine poly (Alizarin red S) modified graphene screen-printed carbon electrode. Curr Nanosci. 2017;13(1):92–9.

    Article  CAS  Google Scholar 

  46. Shinde SM, Kalita G, Sharma S, Papon R, Yusop MZ, Tanemura M. Synthesis of a three dimensional structure of vertically aligned carbon nanotubes and graphene from a single solid carbon source. RSC Adv. 2014;4(26):13355.

    Article  CAS  Google Scholar 

  47. Gupta S, Smith T, Banaszak A, Boeckl J. Graphene quantum dots electrochemistry and sensitive electrocatalytic glucose sensor development. Nanomaterials. 2017;7(10):301.

    Article  PubMed Central  CAS  Google Scholar 

  48. Bonanni A, Chua CK, Pumera M. Rational design of carboxyl groups perpendicularly attached to a graphene sheet: a platform for enhanced biosensing applications. Chem Eur J. 2014;20(1):217–22.

    Article  CAS  PubMed  Google Scholar 

  49. Dadkhah S, Ziaei E, Kayyal TB, Jabbari A, Mehdinia A. A glassy carbon electrode modified with amino-functionalized graphene oxide and molecularly imprinted polymer for electrochemical sensing of bisphenol A. Microchimica Acta (Online). 2016;183(6):1933–41.

    Article  CAS  Google Scholar 

  50. Lu S, Yao M, Yang X, Li Q, Xiao J, Yao Z, et al. High pressure transformation of graphene nanoplates: a Raman study. Chem Phys Lett. 2013;585:101–6.

    Article  CAS  Google Scholar 

  51. Zhang S, Guo S, Chen Z, Wang Y, Gao H, Gomez-Herrero J, et al. Recent progress in 2D group-VA semiconductors: from theory to experiment. Chem Soc Rev. 2018;47(3):982–1021.

    Article  CAS  PubMed  Google Scholar 

  52. Pasinszki T, Krebsz M, Tung TT, Losic D. Carbon nanomaterial based biosensors for non-invasive detection of cancer and disease biomarkers for clinical diagnosis. Sensors. 2017;17(8):1919.

    Article  CAS  PubMed Central  Google Scholar 

  53. Huang B, Liu J, Lai L, Yu F, Ying X, Ye B-C, et al. A free-standing electrochemical sensor based on graphene foam-carbon nanotube composite coupled with gold nanoparticles and its sensing application for electrochemical determination of dopamine and uric acid. J Electroanal Chem. 2017;801:129–34.

    Article  CAS  Google Scholar 

  54. Brownson DA, Kampouris DK, Banks CE. Graphene electrochemistry: fundamental concepts through to prominent applications. Chem Soc Rev. 2012;41(21):6944–76.

    Article  CAS  PubMed  Google Scholar 

  55. Ambrosi A, Chua CK, Bonanni A, Pumera M. Electrochemistry of graphene and related materials. Chem Rev. 2014;114(14):7150–88.

    Article  CAS  PubMed  Google Scholar 

  56. Yuan W, Zhou Y, Li Y, Li C, Peng H, Zhang J, et al. The edge- and basal-plane-specific electrochemistry of a single-layer graphene sheet. Sci Rep. 2013;3:2248.

    Article  PubMed  PubMed Central  Google Scholar 

  57. Bellunato A, Arjmandi Tash H, Cesa Y, Schneider GF. Chemistry at the edge of graphene. Chemphyschem. 2016;17(6):785–801.

    Article  CAS  PubMed  Google Scholar 

  58. Lim CX, Hoh HY, Ang PK, Loh KP. Direct voltammetric detection of DNA and pH sensing on epitaxial graphene: an insight into the role of oxygenated defects. Anal Chem. 2010;82(17):7387–93.

    Article  CAS  PubMed  Google Scholar 

  59. Ferrari AC, Basko DM. Raman spectroscopy as a versatile tool for studying the properties of graphene. Nat Nanotechnol. 2013;8:235.

    Article  CAS  PubMed  Google Scholar 

  60. Wu J-B, Lin M-L, Cong X, Liu H-N, Tan P-H. Raman spectroscopy of graphene-based materials and its applications in related devices. Chem Soc Rev. 2018;47(5):1822–73.

    Article  CAS  PubMed  Google Scholar 

  61. Ferrari AC, Basko DM. Raman spectroscopy as a versatile tool for studying the properties of graphene. Nat Nanotechnol. 2013;8(4):235.

    Article  CAS  PubMed  Google Scholar 

  62. Toh SY, Loh KS, Kamarudin SK, Daud WRW. Graphene production via electrochemical reduction of graphene oxide: synthesis and characterisation. Chem Eng J. 2014;251:422–34.

    Article  CAS  Google Scholar 

  63. Bo X, Zhou M, Guo L. Electrochemical sensors and biosensors based on less aggregated graphene. Biosens Bioelectron. 2017;89:167–86.

    Article  CAS  PubMed  Google Scholar 

  64. Akhavan O, Ghaderi E, Rahighi R. Toward single-DNA electrochemical biosensing by graphene nanowalls. ACS Nano. 2012;6(4):2904–16.

    Article  CAS  PubMed  Google Scholar 

  65. Tian B, Zhou M, Yang Y, Yu L, Luo Z, Tian D, et al. Lab-attenuated rabies virus causes abortive infection and induces cytokine expression in astrocytes by activating mitochondrial antiviral-signaling protein signaling pathway. Front Immunol. 2018;8:2011.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  66. Feicht P, Eigler S. Defects in graphene oxide as structural motifs. ChemNanoMat. 2018;4(3):244–52.

    Article  CAS  Google Scholar 

  67. Ahmadian Yazdi A, Xu J. Nitrogen-doped graphene approach to enhance the performance of a membraneless enzymatic biofuel cell. Front Energy. 2018;12(2):233–8.

    Article  Google Scholar 

  68. Li J, Zeng X, Ren T, van der Heide E. The preparation of graphene oxide and its derivatives and their application in bio-tribological systems. Lubricants. 2014;2(3):137–61.

    Article  Google Scholar 

  69. Botas C, Pérez-Mas AM, Álvarez P, Santamaría R, Granda M, Blanco C, et al. Optimization of the size and yield of graphene oxide sheets in the exfoliation step. Carbon. 2013;63:576–8.

    Article  CAS  Google Scholar 

  70. Qi G-Q, Cao J, Bao R-Y, Liu Z-Y, Yang W, Xie B-H, et al. Tuning the structure of graphene oxide and the properties of poly(vinyl alcohol)/graphene oxide nanocomposites by ultrasonication. J Mater Chem A. 2013;1(9):3163.

    Article  CAS  Google Scholar 

  71. Durge R, Kshirsagar RV, Tambe P. Effect of sonication energy on the yield of graphene nanosheets by liquid-phase exfoliation of graphite. Procedia Eng. 2014;97:1457–65.

    Article  CAS  Google Scholar 

  72. Ye S, Feng J. The effect of sonication treatment of graphene oxide on the mechanical properties of the assembled films. RSC Adv. 2016;6(46):39681–7.

    Article  CAS  Google Scholar 

  73. Zeng F, Sun Z, Sang X, Diamond D, Lau KT, Liu X, et al. In situ one-step electrochemical preparation of graphene oxide nanosheet-modified electrodes for biosensors. Chemsuschem. 2011;4(11):1587–91.

    Article  CAS  PubMed  Google Scholar 

  74. Cui P, Lee J, Hwang E, Lee H. One-pot reduction of graphene oxide at subzero temperatures. Chem Commun. 2011;47(45):12370–2.

    Article  CAS  Google Scholar 

  75. Abid X, Sehrawat P, Islam SS, Mishra P, Ahmad S. Reduced graphene oxide (rGO) based wideband optical sensor and the role of temperature, defect states and quantum efficiency. Sci Rep. 2018;8(1):3537.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  76. Hontoria-Lucas C, López-Peinado AJ, López-González JDD, Rojas-Cervantes ML, Martín-Aranda RM. Study of oxygen-containing groups in a series of graphite oxides: physical and chemical characterization. Carbon. 1995;33(11):1585–92.

    Article  CAS  Google Scholar 

  77. Cho SH, Kwon SS, Yi J, Park WI. Chemical and biological sensors based on defect-engineered graphene mesh field-effect transistors. Nano Conv. 2016;3(1):14.

    Article  CAS  Google Scholar 

  78. Rowley-Neale SJ, Randviir EP, Abo Dena AS, Banks CE. An overview of recent applications of reduced graphene oxide as a basis of electroanalytical sensing platforms. Appl Mater Today. 2018;10:218–26.

    Article  Google Scholar 

  79. Báez D, Pardo H, Laborda I, Marco J, Yáñez C, Bollo S. Reduced graphene oxides: influence of the reduction method on the electrocatalytic effect towards nucleic acid oxidation. Nanomaterials. 2017;7(7):168.

    Article  PubMed Central  CAS  Google Scholar 

  80. Kumar V, Brent JR, Shorie M, Kaur H, Chadha G, Thomas AG, et al. Nanostructured aptamer-functionalized black phosphorus sensing platform for label-free detection of myoglobin, a cardiovascular disease biomarker. ACS Appl Mater Interfaces. 2016;8(35):22860–8.

    Article  CAS  PubMed  Google Scholar 

  81. Tegou E, Pseiropoulos G, Filippidou MK, Chatzandroulis S. Low-temperature thermal reduction of graphene oxide films in ambient atmosphere: infra-red spectroscopic studies and gas sensing applications. Microelectron Eng. 2016;159:146–50.

    Article  CAS  Google Scholar 

  82. Baez DF, Pardo H, Laborda I, Marco JF, Yanez C, Bollo S. Reduced graphene oxides: influence of the reduction method on the electrocatalytic effect towards nucleic acid oxidation. Nanomaterials. 2017;7:7.

    Article  CAS  Google Scholar 

  83. Ferrari AC, Robertson J. Interpretation of Raman spectra of disordered and amorphous carbon. Phys Rev B. 2000;61(20):14095–107.

    Article  CAS  Google Scholar 

  84. Cançado LG, Takai K, Enoki T, Endo M, Kim YA, Mizusaki H, et al. General equation for the determination of the crystallite size La of nanographite by Raman spectroscopy. Appl Phys Lett. 2006;88(16):163106.

    Article  CAS  Google Scholar 

  85. Bousa M, Frank O, Kavan L. Progressive in situ reduction of graphene oxide studied by raman spectroelectrochemistry: implications for a spontaneous activation of LiFePO4 (Olivine). Electroanalysis. 2014;26(1):57–61.

    Article  CAS  Google Scholar 

  86. Isin D, Eksin E, Erdem A. Graphene oxide modified single-use electrodes and their application for voltammetric miRNA analysis. Mater Sci Eng, C. 2017;75:1242–9.

    Article  CAS  Google Scholar 

  87. Lu C, Huang PJ, Liu B, Ying Y, Liu J. Comparison of graphene oxide and reduced graphene oxide for DNA adsorption and sensing. Langmuir. 2016;32(41):10776–83.

    Article  CAS  PubMed  Google Scholar 

  88. Jampasa S, Siangproh W, Duangmal K, Chailapakul O. Electrochemically reduced graphene oxide-modified screen-printed carbon electrodes for a simple and highly sensitive electrochemical detection of synthetic colorants in beverages. Talanta. 2016;160:113–24.

    Article  CAS  PubMed  Google Scholar 

  89. Flox C, Rubio-Garcia J, Nafria R, Zamani R, Skoumal M, Andreu T, et al. Active nano-CuPt3 electrocatalyst supported on graphene for enhancing reactions at the cathode in all-vanadium redox flow batteries. Carbon. 2012;50(6):2372–4.

    Article  CAS  Google Scholar 

  90. Wang C, Du J, Wang H, Zou C, Jiang F, Yang P, et al. A facile electrochemical sensor based on reduced graphene oxide and Au nanoplates modified glassy carbon electrode for simultaneous detection of ascorbic acid, dopamine and uric acid. Sens Actuat. 2014;204:302–9.

    Article  CAS  Google Scholar 

  91. Qian T, Yu C, Zhou X, Wu S, Shen J. Au nanoparticles decorated polypyrrole/reduced graphene oxide hybrid sheets for ultrasensitive dopamine detection. Sens Actuat. 2014;193:759–63.

    Article  CAS  Google Scholar 

  92. Balasubramanian K, Biswas T, Ghosh P, Suran S, Mishra A, Mishra R, et al. Reversible defect engineering in graphene grain boundaries. Nat Commun. 2019;10(1):1090.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  93. Voiry D, Yang J, Kupferberg J, Fullon R, Lee C, Jeong HY, et al. High-quality graphene via microwave reduction of solution-exfoliated graphene oxide. Science. 2016;353(6306):1413–6.

    Article  CAS  PubMed  Google Scholar 

  94. Chowdhury I, Mansukhani ND, Guiney LM, Hersam MC, Bouchard D. Aggregation and stability of reduced graphene oxide: complex roles of divalent cations, pH, and natural organic matter. Environ Sci Technol. 2015;49(18):10886–93.

    Article  CAS  PubMed  Google Scholar 

  95. Gudarzi MM. Colloidal stability of graphene oxide: aggregation in two dimensions. Langmuir. 2016;32(20):5058–68.

    Article  CAS  PubMed  Google Scholar 

  96. Gao J, Yuan Q, Ye C, Guo P, Du S, Lai G, et al. Label-Free electrochemical detection of vanillin through low-defect graphene electrodes modified with Au nanoparticles. Materials. 2018;11:4.

    Google Scholar 

  97. Cui X, Wu S, Li Y, Wan G. Sensing hydrogen peroxide using a glassy carbon electrode modified with in situ electrodeposited platinum-gold bimetallic nanoclusters on a graphene surface. Microchimica Acta (Online). 2015;182(1–2):265–72.

    Article  CAS  Google Scholar 

  98. Wang J, Yang B, Zhong J, Yan B, Zhang K, Zhai C, et al. Dopamine and uric acid electrochemical sensor based on a glassy carbon electrode modified with cubic Pd and reduced graphene oxide nanocomposite. J Colloid Interface Sci. 2017;497:172–80.

    Article  CAS  PubMed  Google Scholar 

  99. Kumar DR, Kesavan S, Baynosa ML, Shim J-J. 3, 5-Diamino-1, 2, 4-triazole@ electrochemically reduced graphene oxide film modified electrode for the electrochemical determination of 4-nitrophenol. Electrochim Acta. 2017;246:1131–40.

    Article  CAS  Google Scholar 

  100. Vasilescu A, Boulahneche S, Chekin F, Gáspár S, Medjram MS, Diagne AA, et al. Porous reduced graphene oxide modified electrodes for the analysis of protein aggregation. Part 1: Lysozyme aggregation at pH 2 and 7.4. Electrochimica Acta. 2017;254:375–83.

    Article  CAS  Google Scholar 

  101. Pourbeyram S, Abdollahpour J, Soltanpour M. Green synthesis of copper oxide nanoparticles decorated reduced graphene oxide for high sensitive detection of glucose. Mater Sci Eng C. 2019;94:850–7.

    Article  CAS  Google Scholar 

  102. Azadmehr F, Zarei K. Ultrasensitive determination of ceftizoxime using pencil graphite electrode modified by hollow gold nanoparticles/reduced graphene oxide. Arabian J Chem. 2018. https://doi.org/10.1016/j.arabjc.2018.02.004.

    Article  Google Scholar 

  103. Manoj D, Theyagarajan K, Saravanakumar D, Senthilkumar S, Thenmozhi K. Aldehyde functionalized ionic liquid on electrochemically reduced graphene oxide as a versatile platform for covalent immobilization of biomolecules and biosensing. Biosens Bioelectron. 2018;103:104–12.

    Article  CAS  PubMed  Google Scholar 

  104. Rajesh SS, Kotnala RK. Single frequency impedance analysis on reduced graphene oxide screen-printed electrode for biomolecular detection. Appl Biochem Biotechnol. 2017;183(2):672–83.

    Article  CAS  PubMed  Google Scholar 

  105. Li S, Zhang Q, Lu Y, Ji D, Zhang D, Wu J, et al. One step electrochemical deposition and reduction of graphene oxide on screen printed electrodes for impedance detection of glucose. Sens Actuat Chem. 2017;244:290–8.

    Article  CAS  Google Scholar 

  106. Zhou M, Zhai Y, Dong S. Electrochemical sensing and biosensing platform based on chemically reduced graphene oxide. Anal Chem. 2009;81(14):5603–13.

    Article  CAS  PubMed  Google Scholar 

  107. Loo AH, Bonanni A, Pumera M. Impedimetric thrombin aptasensor based on chemically modified graphenes. Nanoscale. 2012;4(1):143–7.

    Article  CAS  PubMed  Google Scholar 

  108. Giovanni M, Bonanni A, Pumera M. Detection of DNA hybridization on chemically modified graphene platforms. Analyst. 2012;137(3):580–3.

    Article  CAS  PubMed  Google Scholar 

  109. Lonkar SP, Deshmukh YS, Abdala AA. Recent advances in chemical modifications of graphene. Nano Res. 2015;8(4):1039–74.

    Article  CAS  Google Scholar 

  110. Chen J, Wang X, Chen T. Facile and green reduction of covalently PEGylated nanographene oxide via a ‘water-only’ route for high-efficiency photothermal therapy. Nanoscale Res Lett. 2014;9(1):86.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  111. Zhang W, Li Z, Gu H, Li Y, Zhang G, Zhang F, et al. l-proline covalently anchored on graphene oxide as an effective bifunctional catalyst for ketene forming reaction. Chem Eng Sci. 2015;135:187–92.

    Article  CAS  Google Scholar 

  112. Bo X, Zhou M, Guo L. Electrochemical sensors and biosensors based on less aggregated graphene. Biosens Bioelectron. 2017;89(Pt 1):167–86.

    Article  CAS  PubMed  Google Scholar 

  113. Mandli J, Mohammadi H, Amine A. Electrochemical DNA sandwich biosensor based on enzyme amplified microRNA-21 detection and gold nanoparticles. Bioelectrochemistry. 2017;116:17–23.

    Article  CAS  PubMed  Google Scholar 

  114. Zhong J-H, Zhang J, Jin X, Liu J-Y, Li Q, Li M-H, et al. Quantitative correlation between defect density and heterogeneous electron transfer rate of single layer graphene. J Am Chem Soc. 2014;136(47):16609–17.

    Article  CAS  PubMed  Google Scholar 

  115. Yuan Q, Liu Y, Ye C, Sun H, Dai D, Wei Q, et al. Highly stable and regenerative graphene–diamond hybrid electrochemical biosensor for fouling target dopamine detection. Biosens Bioelectron. 2018;111:117–23.

    Article  CAS  PubMed  Google Scholar 

  116. Ye Y, Xie J, Ye Y, Cao X, Zheng H, Xu X, et al. A label-free electrochemical DNA biosensor based on thionine functionalized reduced graphene oxide. Carbon. 2018;129:730–7.

    Article  CAS  Google Scholar 

  117. Xuan X, Yoon HS, Park JY. A wearable electrochemical glucose sensor based on simple and low-cost fabrication supported micro-patterned reduced graphene oxide nanocomposite electrode on flexible substrate. Biosens Bioelectron. 2018;109:75–82.

    Article  CAS  PubMed  Google Scholar 

  118. Zhu X, Xu J, Duan X, Lu L, Zhang K, Yu Y, et al. Controlled synthesis of partially reduced graphene oxide: enhance electrochemical determination of isoniazid with high sensitivity and stability. J Electroanal Chem. 2015;757:183–91.

    Article  CAS  Google Scholar 

  119. Rhodes D, Chae SH, Ribeiro-Palau R, Hone J. Disorder in van der Waals heterostructures of 2D materials. Nat Mater. 2019;18(6):541.

    Article  CAS  PubMed  Google Scholar 

  120. Timm RA, Kisner A, Bassetto VC, Kubota LT. Critical view on graphene oxide production and its transfer to surfaces aiming electrochemical applications. J Nanosci Nanotechnol. 2014;14(9):6478–96.

    Article  CAS  PubMed  Google Scholar 

  121. Liu Y, Gao L, Sun J, Wang Y, Zhang J. Stable Nafion-functionalized graphene dispersions for transparent conducting films. Nanotechnology. 2009;20(46):465605.

    Article  PubMed  CAS  Google Scholar 

  122. Yang F, Liu Y, Gao L, Sun J. pH-Sensitive highly dispersed reduced graphene oxide solution using lysozyme via an in situ reduction method. J Phys Chem. 2010;114(50):22085–91.

    CAS  Google Scholar 

  123. Zhao Y, Zhang YH, Zhuge Z, Tang YH, Tao JW, Chen Y. Synthesis of a poly-l-lysine/black phosphorus hybrid for biosensors. Anal Chem. 2018;90(5):3149–55.

    Article  CAS  PubMed  Google Scholar 

  124. Bollella P, Fusco G, Tortolini C, Sanzo G, Favero G, Gorton L, et al. Beyond graphene: electrochemical sensors and biosensors for biomarkers detection. Biosens Bioelectron. 2017;89(Pt 1):152–66.

    Article  CAS  PubMed  Google Scholar 

  125. Lv W, Yang B, Wang B, Wan W, Ge Y, Yang R, et al. Sulfur-doped black phosphorus field-effect transistors with enhanced stability. ACS Appl Mater Interfaces. 2018;10(11):9663–8.

    Article  CAS  PubMed  Google Scholar 

  126. Campos R, Machado G Jr, Cerqueira M, Borme J, Alpuim P. Wafer scale fabrication of graphene microelectrode arrays for the detection of DNA hybridization. Microelectron Eng. 2018;189:85–90.

    Article  CAS  Google Scholar 

  127. Bonanni A, Pumera M. Graphene platform for hairpin-DNA-based impedimetric genosensing. ACS Nano. 2011;5(3):2356–61.

    Article  CAS  PubMed  Google Scholar 

  128. Georgakilas V, Otyepka M, Bourlinos AB, Chandra V, Kim N, Kemp KC, et al. Functionalization of graphene: covalent and non-covalent approaches, derivatives and applications. Chem Rev. 2012;112(11):6156–214.

    Article  CAS  PubMed  Google Scholar 

  129. Benvidi A, Rajabzadeh N, Zahedi HM, Mazloum-Ardakani M, Heidari MM, Hosseinzadeh L. Simple and label-free detection of DNA hybridization on a modified graphene nanosheets electrode. Talanta. 2015;137:80–6.

    Article  CAS  PubMed  Google Scholar 

  130. Lin L, Peng H, Liu Z. Synthesis challenges for graphene industry. Nat Mater. 2019;18(6):520.

    Article  CAS  PubMed  Google Scholar 

  131. Thangamuthu M, Hsieh KY, Kumar PV, Chen G-Y. Graphene-and graphene oxide-based nanocomposite platforms for electrochemical biosensing applications. Int J Mol Sci. 2019;20(12):2975.

    Article  PubMed Central  Google Scholar 

  132. Lee H, Paeng K, Kim IS. A review of doping modulation in graphene. Synth Met. 2018;244:36–47.

    Article  CAS  Google Scholar 

  133. Primiceri E, Chiriacò M, Notarangelo F, Crocamo A, Ardissino D, Cereda M, et al. Key enabling technologies for point-of-Care diagnostics. Sensors. 2018;18(11):3607.

    Article  CAS  PubMed Central  Google Scholar 

  134. Lee J-H, Park S-J, Choi J-W. Electrical property of graphene and its application to electrochemical biosensing. Nanomaterials. 2019;9(2):297.

    Article  CAS  PubMed Central  Google Scholar 

  135. Li B, Pan G, Avent ND, Lowry RB, Madgett TE, Waines PL. Graphene electrode modified with electrochemically reduced graphene oxide for label-free DNA detection. Biosens Bioelectron. 2015;72:313–9.

    Article  CAS  PubMed  Google Scholar 

  136. Bonanni A, Chua CK, Pumera M. Rational design of carboxyl groups perpendicularly attached to a graphene sheet: a platform for enhanced biosensing applications. Chemistry. 2014;20(1):217–22.

    Article  CAS  PubMed  Google Scholar 

  137. Dubuisson E, Yang Z, Loh KP. Optimizing label-free DNA electrical detection on graphene platform. Anal Chem. 2011;83(7):2452–60.

    Article  CAS  PubMed  Google Scholar 

  138. Wang Y, Sauriat-Dorizon H, Korri-Youssoufi H. Direct electrochemical DNA biosensor based on reduced graphene oxide and metalloporphyrin nanocomposite. Sens Actuat Chemical. 2017;251:40–8.

    Article  CAS  Google Scholar 

  139. Wang L, Qin X, Liu S, Luo Y, Asiri AM, Al-Youbi AO, et al. Single-stranded DNA-mediated immobilization of graphene on a gold electrode for sensitive and selective determination of dopamine. ChemPlusChem. 2012;77(1):19–22.

    Article  CAS  Google Scholar 

  140. Chu Y, Cai B, Ma Y, Zhao M, Ye Z, Huang J. Highly sensitive electrochemical detection of circulating tumor DNA based on thin-layer MoS2/graphene composites. RSC Adv. 2016;6(27):22673–8.

    Article  CAS  Google Scholar 

  141. Fang L-X, Cao J-T, Huang K-J. A sensitive electrochemical biosensor for specific DNA sequence detection based on flower-like VS2, graphene and Au nanoparticles signal amplification. J Electroanal Chem. 2015;746:1–8.

    Article  CAS  Google Scholar 

  142. Muti M, Sharma S, Erdem A, Papakonstantinou P. Electrochemical monitoring of nucleic acid hybridization by single-use graphene oxide-based sensor. Electroanalysis. 2011;23(1):272–9.

    Article  CAS  Google Scholar 

  143. Zhao J, Chen G, Zhu L, Li G. Graphene quantum dots-based platform for the fabrication of electrochemical biosensors. Electrochem Commun. 2011;13(1):31–3.

    Article  CAS  Google Scholar 

  144. Wang Z, Zhang J, Chen P, Zhou X, Yang Y, Wu S, et al. Label-free, electrochemical detection of methicillin-resistant Staphylococcus aureus DNA with reduced graphene oxide-modified electrodes. Biosens Bioelectron. 2011;26(9):3881–6.

    Article  CAS  PubMed  Google Scholar 

  145. Feng L, Chen Y, Ren J, Qu X. A graphene functionalized electrochemical aptasensor for selective label-free detection of cancer cells. Biomaterials. 2011;32(11):2930–7.

    Article  CAS  PubMed  Google Scholar 

  146. Banhart F, Kotakoski J, Krasheninnikov AV. Structural defects in graphene. ACS Nano. 2011;5(1):26–41.

    Article  CAS  PubMed  Google Scholar 

  147. Huang PY, Ruiz-Vargas CS, Van Der Zande AM, Whitney WS, Levendorf MP, Kevek JW, et al. Grains and grain boundaries in single-layer graphene atomic patchwork quilts. Nature. 2011;469(7330):389.

    Article  CAS  PubMed  Google Scholar 

  148. Wang Y, Shao Y, Matson DW, Li J, Lin Y. Nitrogen-doped graphene and its application in electrochemical biosensing. ACS Nano. 2010;4(4):1790–8.

    Article  CAS  PubMed  Google Scholar 

  149. Cançado LG, Da Silva MG, Ferreira EHM, Hof F, Kampioti K, Huang K, et al. Disentangling contributions of point and line defects in the Raman spectra of graphene-related materials. Materials. 2017;4(2):025039.

    Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was supported by a grant of the Ministry of Research and Innovation, Operational Program Competitive Axis 1—Section E, Program co-financed from European Regional Development Fund “Investments for your future” under the project number 154/25.11.2016, P_37_221/2015, “A novel graphene biosensor testing osteogenic potency; capturing best stem cell performance for regenerative medicine” (GRABTOP).

Author information

Authors and Affiliations

Authors

Contributions

TT conceived ideas, drafted the manuscript and selected figures. EAC compiled tables, contributed to figure and text revision. JSB reviewed concepts, revised and edited the manuscript. MI reviewed concepts, revised and edited the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Mariana Ioniţă.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tite, T., Chiticaru, E.A., Burns, J.S. et al. Impact of nano-morphology, lattice defects and conductivity on the performance of graphene based electrochemical biosensors. J Nanobiotechnol 17, 101 (2019). https://doi.org/10.1186/s12951-019-0535-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12951-019-0535-6

Keywords